USER'S MANUAL<br />

General instructions for use<br />

6.8.2.2 <strong>Manual</strong> exposure<br />

If the kV, mA and s combinations of the above table are not considered<br />

suitable for a specific examination, it will be possible to set new<br />

parameters using the manual mode.<br />

To change the kV or mA values, press key "7" , or key "5" .<br />

The LED on the Patient Size key switches off and the Adult / Child and<br />

Cephalometry keys will flash. One of the following messages will be<br />

displayed:<br />

8 x 1 0 A S T F = 0 8 . 3<br />

> 7 0 k V 1 0 m A 0 . 8 0 s<br />

8 x 1 0 A S T F = 0 8 . 3<br />

7 0 k V > 1 0 m A 0 . 8 0 s<br />

8 x 1 0 A S T F = 0 8 . 3<br />

7 0 k V 1 0 m A > 0 . 8 0 s<br />

The symbol ">" indicates which parameter is being changed.<br />

The selected parameter can be changed using the increase key "3"<br />

and decrease key "6" .<br />

The "kV" value can vary between 60 and 80 kV, with 2 kV steps.<br />

The "mA" value can vary between 4 and 12 mA, with 1 mA steps.<br />

The "s" value can vary between 0.20 s and 3.00 s as per the following<br />

table:<br />

Exposure times selectable in manual mode<br />

0,20 - 0,23 - 0,26 - 0,30 - 0,33 - 0,36 - 0,40 - 0,45 - 0,50 - 0,60<br />

- 0,70 - 0,80- 0,90- 1,00 - 1,30- 1,60 - 2,00 - 3,00<br />

Table 7<br />

<br />

NOTE:<br />

To change the values rapidly, keep the increase key "3"<br />

decrease key "6"<br />

pressed.<br />

or<br />

STRATO X (120V) 110<br />

(Rev. 1)