USER'S MANUAL<br />
General instructions for use<br />
6.8.2.2 <strong>Manual</strong> exposure<br />
If the kV, mA and s combinations of the above table are not considered<br />
suitable for a specific examination, it will be possible to set new<br />
parameters using the manual mode.<br />
To change the kV or mA values, press key "7" , or key "5" .<br />
The LED on the Patient Size key switches off and the Adult / Child and<br />
Cephalometry keys will flash. One of the following messages will be<br />
displayed:<br />
8 x 1 0 A S T F = 0 8 . 3<br />
> 7 0 k V 1 0 m A 0 . 8 0 s<br />
8 x 1 0 A S T F = 0 8 . 3<br />
7 0 k V > 1 0 m A 0 . 8 0 s<br />
8 x 1 0 A S T F = 0 8 . 3<br />
7 0 k V 1 0 m A > 0 . 8 0 s<br />
The symbol ">" indicates which parameter is being changed.<br />
The selected parameter can be changed using the increase key "3"<br />
and decrease key "6" .<br />
The "kV" value can vary between 60 and 80 kV, with 2 kV steps.<br />
The "mA" value can vary between 4 and 12 mA, with 1 mA steps.<br />
The "s" value can vary between 0.20 s and 3.00 s as per the following<br />
table:<br />
Exposure times selectable in manual mode<br />
0,20 - 0,23 - 0,26 - 0,30 - 0,33 - 0,36 - 0,40 - 0,45 - 0,50 - 0,60<br />
- 0,70 - 0,80- 0,90- 1,00 - 1,30- 1,60 - 2,00 - 3,00<br />
Table 7<br />