7. At this point, the patient must move his feet towards the keyboard. In<br />
this way, a better straightening of the spine in the cervical area is<br />
obtained thus improving the darkening of the radiography.<br />
8. Also, close the temple support acting on the two knobs "6" (Figure 34<br />
and Figure 35). The temple support will help the patient to stay in a<br />
correct position. Check that, during this phase, patient has not<br />
changed the position.<br />
<br />
NOTE:<br />
The laser centring device remains lit until the button "20"<br />
is<br />
pressed again or, upon completion of the alignment procedure, button<br />
"23" is pressed to start exposure.<br />
9. Once alignment has been completed, the following message will be<br />
displayed:<br />
S T A R T E X A M<br />
x x k V x x m A 1 5 . 0 s<br />
x = value defined by the settings<br />
10. Make the following recommendations to the patient: the mouth<br />
must remain closed, he/she must remain perfectly still and do<br />
not look at the rotating arm during movements.<br />
STRATO X (120V) 98<br />
(Rev. 1)