USER'S MANUAL<br />
General instructions for use<br />
6.6.4 Making an exposure<br />
<br />
WARNING:<br />
During the emission of X-rays, the protection procedures for the operator<br />
and personnel in the area must be in compliance with the local<br />
regulations.<br />
In all cases, it is recommended that during the emission of X-rays, only<br />
the patient and operator be present in the room. If the operator is not<br />
protected by suitable screens, he must stand at least 1.5 meters away<br />
from the emission of the X-rays (see the Figure 1 and Figure 2).<br />
<br />
NOTE:<br />
If deemed necessary it is possible to check the interference of the rotation<br />
movement with the shoulder of the patient by activating the Test<br />
function, pressing key "8" .<br />
In this condition, it will be possible to make the unit perform all the<br />
movements made during the examination without emitting X-rays. The<br />
test function of the Biaxial TMJ is the same as for panoramic mode.<br />
Once the cycle has been completed, deactivate the "Test" function and<br />
press key "8" again.<br />
1. Verify once again that the exposure data are correct, if not, correct<br />
them as described in paragraph 6.4.2.2. Check that the unit ready<br />
for X-ray light "1" is on, then press X-ray button "30" for the entire<br />
duration of the exposure, ensuring that the X-ray light "2" is also lit<br />
(if the unit is in view) and that the X-ray signal sounds. The following<br />
message will be displayed first:<br />
E X A M<br />
P R E - H E A T I N G<br />
and then (after 2 seconds), the following message will be displayed:<br />
> X - R A Y <<br />
x x k V x x m A x . x x s<br />
x = the value defined by the settings<br />
<br />
NOTE:<br />
If there is no cassette, the following message will be displayed:<br />
> S T A R T E X A M <<br />
I N S E R T C A S S E T T E<br />
The examination will not be continued until the cassette has been<br />
inserted.<br />
STRATO X (120V) 72<br />
(Rev. 1)