USER'S MANUAL<br />
General instructions for use<br />
7. At this point, the patient must move his feet towards the keyboard<br />
making sure to maintain the head within the above mentioned<br />
pre-aligned anatomical references. In this way, a greater<br />
straightening of the spine in the cervical area is obtained and<br />
collision between the cassette holder and the patient's shoulder is<br />
prevented. Check that the Frankfurt plane is still horizontal.<br />
8. Also, close the temple support acting on the two knobs "6" (Figure<br />
12). The temple support will help the patient to stay in a correct<br />
position. Check that, during this phase, patient has not changed the<br />
position.<br />
9. Press key "23" to confirm. The confirmation request LED<br />
switches off. The luminous centering devices switch off and the<br />
rotating arm goes to its examination start position. Once alignment<br />
has been completed, the following message will be displayed:<br />
S T A R T E X A M<br />
x x k V x x m A 1 5 . 0 s<br />
x = value defined by the settings<br />
<br />
NOTE:<br />
The laser centering device remains lit until the button "20"<br />
pressed again or, upon completion of the alignment procedure, key<br />
is<br />
"23" is pressed to start the exam procedure.<br />
10. Make the following recommendations to the patient: the lips<br />
must remain closed, the tongue must be placed towards the<br />
palate, he/she must remain perfectly still and do not look at the<br />
rotating arm during the movements.<br />
(Rev. 1)<br />
45<br />
STRATO X (120V)