General instructions for use<br />
<br />
NOTE:<br />
If there is no cassette, the following message will be displayed:<br />
C E P H<br />
I N S E R T F I L M .<br />
<br />
NOTE:<br />
If, during the preparation of the first cephalometric examination the<br />
system detects a cassette which has not been replaced, the following<br />
message will be displayed:<br />
C E P H<br />
R E M O V E F I L M<br />
The examination will not be continued until a new cassette has been<br />
inserted.<br />
<br />
NOTE:<br />
X-rays are emitted with a delay of two seconds from pressing the X-ray<br />
button to allow the heating of the filament and the control of all set<br />
parameters. Since the X-ray button is a "dead man" control, it must<br />
be kept pressed until the end of the exposure.<br />
3. Once the exposure is completed, the following message will be<br />
displayed:<br />
C E P H<br />
R E M O V E F I L M<br />
4. Remove the cassette. LED "23" lights up and the set cassette,<br />
STF, kV, mA and seconds values relative to the exposure made will<br />
be displayed.<br />
<br />
NOTE:<br />
If you try to perform a new exam before the cooling period has elapse (4<br />
minutes), the following message will be displayed indicating the time to<br />
wait before performing a new examination:<br />
T U B E C O O L I N G<br />
P L E A S E W A I T x x x s<br />
This time enables the X-ray tube's anode to cool down.<br />