26.01.2014 Views

Strato X User Manual - Image Works

Strato X User Manual - Image Works

Strato X User Manual - Image Works

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

USER'S MANUAL<br />

General instructions for use<br />

<br />

WARNING:<br />

The position of the identification characters "R" (right side) and "L" (left<br />

side) are correct if the cassette is fit into the unit with the hinges<br />

positioned upward.<br />

<br />

NOTE:<br />

The rotation of the arm and the emission of the X-rays will start with a<br />

delay of 2 seconds from pressing the X-ray button "30". Since the X-ray<br />

button is a "dead man" control, it must be kept pressed until the end<br />

of the exposure.<br />

2. Once the exposure is completed, the system will rotate back to a<br />

position where it is possible to make the patient to exit. At the end of<br />

this movement, the following message will be displayed:<br />

P A T I E N T E X I T<br />

P R E S S → 0 ←<br />

At this point, the patient can be freed from the positioning device.<br />

<br />

NOTE:<br />

If the examination is made in "Test" mode with the patient already in<br />

position, he must not be removed from the temple support group to avoid<br />

having to reposition him.<br />

Press button "10"<br />

to return the unit to its initial position. This<br />

movement can be stopped by pressing button "9" .<br />

Now the system is ready to perform a new exam.<br />

3. Press button "10" . The unit will return to its initial position<br />

and the following message will be displayed:<br />

P L E A S E W A I T . . .<br />

<br />

WARNING:<br />

During this phase the system stores into the internal memory some data.<br />

Switching off the unit during this phase can lead to the loss of some<br />

set-up data normally stored in the memory, which in turn may cause<br />

irreversible damages to the system itself.<br />

Do not switch off the unit in this phase.<br />

If required, switch off the unit only when the green LED "23"<br />

is lit.<br />

(Rev. 1)<br />

47<br />

STRATO X (120V)