General instructions for use<br />

6.5 TMJ examination<br />

The TMJ examination with open or closed mouth is similar to panoramic,<br />

the only difference is that the exposure is performed only on the involved<br />

area (Temporal mandibular Joint). then it stops and starts on the second<br />

articulation. The operation sequence of the examination is therefore<br />

similar to the one described for the panoramic.<br />

This TMJ function enables to obtain 4 different images on the same film,<br />

by performing two rotational movements. The 4 images represent the<br />

right and left condyle of the temporal-mandibular arch (TMJ) with closed<br />

mouth and open mouth.<br />

Selecting closed mouth exam only the external sectors of the film are<br />

exposed, while selecting open mouth exam, the exposure occurs on the<br />

inner sectors.<br />

The position of the images couples the images corresponding to the same<br />

condyle to help a diagnosis. The Figure 14 shows the information related<br />

to the single sectors.<br />

RIGHT condyle<br />

with closed mouth<br />

RIGHT condyle<br />

with open mouth<br />

LEFT condyle<br />

with open mouth<br />

LEFT condyle<br />

with closed mouth<br />

1 st exposure<br />

3 rd exposure<br />

4 th exposure<br />

2 nd exposure<br />

R<br />

L<br />

Figure 14<br />

<br />

NOTE:<br />

During TMJ the X-rays emission is not continuous since it is interrupted<br />

during the transition phases between the different exposures, but is<br />

necessary to keep pressed the X-ray button "30" for the whole rotation<br />

time.<br />

Do not release the X-ray button during the emission interruption if not<br />

necessary.<br />

The cooling phase of the tubehead occurs at the end of all 4 exposures.<br />

In the CHILD position, exposure start is delayed by a few degrees with<br />

respect to the ADULT position.<br />