Strato X User Manual - Image Works
Strato X User Manual - Image Works
Strato X User Manual - Image Works
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
USER'S MANUAL<br />
General instructions for use<br />
6.7.7 Making an exposure<br />
<br />
NOTE:<br />
When key "8"<br />
is pressed, the Test function will be activated. In<br />
this condition, it will be possible to make the unit perform all the<br />
movements made during the examination without emitting X-rays.<br />
Once the cycle has been completed, deactivate the "Test" by pressing key<br />
"8" again.<br />
<br />
WARNING:<br />
During the emission of X-rays, the protection procedures for the operator<br />
and personnel in the area must be in compliance with the regulations in<br />
force in the country where the machine is used.<br />
In all cases, it is recommended that during the emission of X-rays, only<br />
the patient and operator be present in the room. If the operator is not<br />
protected by suitable screens, he must stand at least 1.5 meters away<br />
from the focal spot of the tubehead.<br />
1. Press button "23" ; display will show:<br />
S T A R T E X A M<br />
x x k V x x m A 1 5 . 0 s<br />
Verify once again that the exposure data are correct, if not, correct<br />
them as described in paragraph 6.7.4. Check that light unit ready for<br />
X-ray is ON, then press X-ray button "30" for the entire duration of<br />
the exposure, ensuring that the X-ray light 2 is also lit (if the unit is<br />
in view) and that the X-ray acoustic signal sounds. The following<br />
message will be displayed first:<br />
> S T A R T E X A M<br />
P R E - H E A T I N G<br />
and then (after 2 seconds), the following message will be displayed:<br />
> X - R A Y <<br />
x x k V x x m A Y . Y Y s<br />
x = the value defined by the settings<br />
during emission, while during intermediate movements the message<br />
will be displayed.<br />
> X - R A Y <<br />
x x k V x x m A Y . Y Y s<br />
x = the value defined by the settings<br />
(Rev. 1)<br />
99<br />
STRATO X (120V)