USER'S MANUAL<br />

General instructions for use<br />

6.7.2.2 Bite block preparing: Mandible Implant<br />

As already stated, this kind of examination has to be performed having<br />

the patient’s head inclined in order to maintain the mandible border in<br />

the point of interest most horizontal as possible; in this way radiographs<br />

will be of better quality and with all clinical interesting points.<br />

1. Remove the standards bite block and insert the special implant bite<br />

block for mandible (main characteristics: can be inclined).<br />

2. Insert the plastic bite, after its cold disaffection, in the bite block<br />

base (please refer to item 5 in Figure 33).<br />

3. Select the correct angle; according to the patient’s morphology: it<br />

must be evaluated case by case. As suggestion, for all teeth it advised<br />

to use the lower angle (refer to Figure 33A). It is possible, however for<br />

some patient, depending on his morphology that the higher angle has<br />

to be used. (please refer to Figure 33C) for posterior teeth. To do this,<br />

referring to Figure 33B:<br />

• unscrew the stopping mechanism using the knob 3<br />

• place the bite block support 4 on the chosen angle<br />

• hold the bite support on the position using the knob 3, in order<br />

to hold the correct position during the examination.<br />

<br />

NOTE:<br />

We suggest to use, as a reference to check the horizontality of the<br />

mandibular border, the laser beam normally used for the Frankfurt<br />

plane. Then choose the correct angle.<br />

4. Take care that the plastic bite is on the higher position, in order to<br />

assure that the area of interest will be irradiated (please refer to the<br />

Figure 33A). To do this, unscrew the stopping mechanism using the<br />

knob 1, place the bite 2 in the highest position and fix the position<br />

using the knob 1, in order to hold the bite in the correct position<br />

during the examination.<br />

Legend:<br />

1 - Knob used to<br />

fix the bite<br />

2 - Bite<br />

3 - Knob to fix the<br />

angulation<br />

4 - Bite block<br />

5 - Base<br />

Figure 33<br />

STRATO X (120V) 90<br />

(Rev. 1)