USER'S MANUAL<br />
General instructions for use<br />
6.7.2.2 Bite block preparing: Mandible Implant<br />
As already stated, this kind of examination has to be performed having<br />
the patient’s head inclined in order to maintain the mandible border in<br />
the point of interest most horizontal as possible; in this way radiographs<br />
will be of better quality and with all clinical interesting points.<br />
1. Remove the standards bite block and insert the special implant bite<br />
block for mandible (main characteristics: can be inclined).<br />
2. Insert the plastic bite, after its cold disaffection, in the bite block<br />
base (please refer to item 5 in Figure 33).<br />
3. Select the correct angle; according to the patient’s morphology: it<br />
must be evaluated case by case. As suggestion, for all teeth it advised<br />
to use the lower angle (refer to Figure 33A). It is possible, however for<br />
some patient, depending on his morphology that the higher angle has<br />
to be used. (please refer to Figure 33C) for posterior teeth. To do this,<br />
referring to Figure 33B:<br />
• unscrew the stopping mechanism using the knob 3<br />
• place the bite block support 4 on the chosen angle<br />
• hold the bite support on the position using the knob 3, in order<br />
to hold the correct position during the examination.<br />
<br />
NOTE:<br />
We suggest to use, as a reference to check the horizontality of the<br />
mandibular border, the laser beam normally used for the Frankfurt<br />
plane. Then choose the correct angle.<br />
4. Take care that the plastic bite is on the higher position, in order to<br />
assure that the area of interest will be irradiated (please refer to the<br />
Figure 33A). To do this, unscrew the stopping mechanism using the<br />
knob 1, place the bite 2 in the highest position and fix the position<br />
using the knob 1, in order to hold the bite in the correct position<br />
during the examination.<br />
Legend:<br />
1 - Knob used to<br />
fix the bite<br />
2 - Bite<br />
3 - Knob to fix the<br />
angulation<br />
4 - Bite block<br />
5 - Base<br />
Figure 33<br />
STRATO X (120V) 90<br />
(Rev. 1)