6.12 Storing of automatical exposure parameters<br />

At the end of mechanical installation and functional check the stored<br />

exposure parameters can be modified depending on the operator needs<br />

such as the combinations between screens and films or the need to get<br />

less or more contrasted (exposed) images.<br />

6.12.1 Exposure parameter<br />

To modify the automatic exposure parameters, please follow the indicated<br />

procedure:<br />

1. Select the exam, the type of patient and the size to be modified.<br />

2. According to the parameter to be modified press "7" and/or<br />

"5" and/or "6" (the latter only for Cephalometric<br />

function). On the display, on the left of the selected parameter value,<br />

a reference character ">" will be displayed.<br />

<br />

NOTE:<br />

The pressure of one of the three parameters setting keys will cause the<br />

de-activation of the patient size LED and the blinking of the patient type<br />

and function/examination LED's. This LED's will be reactivated only<br />

when storing parameters (paragraph 6.12.3).<br />

3. Press "3" or "6" to set the value to store.<br />

STRATO X (120V) 138<br />