26.01.2014 Views

Strato X User Manual - Image Works

Strato X User Manual - Image Works

Strato X User Manual - Image Works

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

USER'S MANUAL<br />

General instructions for use<br />

6.4.4 Making an exposure<br />

<br />

NOTE:<br />

When key "8"<br />

is pressed, the Test function will be activated. In<br />

this condition, it will be possible to make the unit perform all the<br />

movements made during the examination without emitting X-rays.<br />

Once the cycle has been completed, deactivate the "Test" by pressing key<br />

"8" again<br />

<br />

WARNING:<br />

During the emission of X-rays, the protection procedures for the operator<br />

and personnel in the area must be in compliance with the local<br />

regulations.<br />

In all cases, it is recommended that during the emission of X-rays, only<br />

the patient and operator be present in the room. If the operator is not<br />

protected by suitable screens, he must stand at least 1.5 meters away<br />

from the emission of the X-rays (see Figure 1 and Figure 2).<br />

1. Verify once again that the exposure data are correct, if not, correct<br />

them as described in paragraph 6.4.2.2. Check that light unit ready<br />

for X-ray "1" is ON, then press X-ray button "30" for the entire<br />

duration of the exposure, ensuring that the X-ray light "2" is also lit<br />

(if the unit is in view) and that the X-ray acoustic signal sounds. The<br />

following message will be displayed first:<br />

E X A M<br />

P R E - H E A T I N G<br />

and then (after 2 seconds), the following message will be displayed:<br />

> X - R A Y <<br />

x x k V x x m A 1 5 . 0 s<br />

x = the value defined by the settings<br />

<br />

NOTE:<br />

If there is no cassette, the following message will be displayed:<br />

> S T A R T E X A M <<br />

I N S E R T C A S S E T T E<br />

The examination will not be continued until the cassette has been<br />

inserted.<br />

STRATO X (120V) 46<br />

(Rev. 1)