USER'S MANUAL<br />

General instructions for use<br />

6.5.3.2 First exposure (Closed mouth)<br />

<br />

WARNING:<br />

During the emission of X-rays, the protection procedures for the operator<br />

and personnel in the area must be in compliance with the local<br />

regulations.<br />

In all cases, it is recommended that during the emission of X-rays, only<br />

the patient and operator be present in the room. If the operator is not<br />

protected by suitable screens, he must stand at least 1.5 meters away<br />

from the emission of the X-rays (see the Figure 1 and Figure 2).<br />

<br />

NOTE:<br />

If deemed necessary it is possible to check the interference of the rotation<br />

movement with the shoulder of the patient by activating the Test<br />

function, pressing key "8" .<br />

In this condition, it will be possible to make the unit perform all the<br />

movements made during the examination without emitting X-rays. The<br />

test function of the TMJ open/closed mouth is the same as for panoramic<br />

mode and so there will not be a second rotation corresponding to the<br />

open mouth exam.<br />

Once the cycle has been completed, deactivate the "Test" function and<br />

press key "8" again.<br />

1. Check once again that the exposure data are correct, if not, correct<br />

them as described in paragraph 6.5.2.2. Check that the unit ready<br />

for X-ray light "1" is on, then press X-ray button "30" for the entire<br />

duration of the exposure, ensuring that the X-ray light "2" is also lit<br />

(if the unit is in view) and that the X-ray signal sounds. The following<br />

message will be displayed first:<br />

E X A M<br />

P R E - H E A T I N G<br />

and then (after 2 seconds), the following message will be displayed:<br />

> X - R A Y <<br />

x x k V x x m A x . x x s<br />

x = the value defined by the settings<br />

<br />

NOTE:<br />

If there is no cassette, the following message will be displayed:<br />

> S T A R T E X A M <<br />

I N S E R T C A S S E T T E<br />

The examination will not be continued until the cassette has been<br />

inserted.<br />