USER'S MANUAL<br />
General instructions for use<br />
6.7.5 Selection of anatomic and tomographic parameters<br />
1. Select the teeth using button "26" or "24" ; in this way<br />
it is possible to select teeth changing the second digit between 1<br />
(scissors) and 8 (third molar). Acting on buttons "22"<br />
or<br />
"25" , it is possible to change the first digit ranging from 1<br />
(right maxilla) to 4 (left mandible). Please note that in anatomic mode<br />
kV and mA changes accordingly to the selected teeth.<br />
The setting is confirmed by pressing button "23" .<br />
2. Display will visualise:<br />
→ ← L O N G I T : 6 m m<br />
5 8 k V 0 7 m A 0 3 . 5 s<br />
Set the longitudinal slice thickness in the range of values of 4, 6 or<br />
8 mm using buttons "26" or "24" .<br />
<br />
NOTE:<br />
In case that values of kV or mA are considered not adequate, both for<br />
longitudinal and/or cross image, it is possible to select a new exposure<br />
parameter’ value using the method of manual exposure as described on<br />
paragraph 6.7.4.<br />
The selection has to be confirmed using button "23" .<br />
3. Display will visualise the message:<br />
→ ← T R A N S V : 4 m m<br />
5 8 k V 0 7 m A 0 3 . 5 s<br />
Set the longitudinal slice thickness between values of 2, 3 or 4mm<br />
using buttons "26" or "24" .<br />
<br />
NOTE:<br />
In case that values of kV or mA are considered not adequate, both for<br />
longitudinal and/or cross image, it is possible to select a new exposure<br />
parameter’ value using the method of manual exposure as described on<br />
paragraph 6.7.4.<br />
Set the chosen value using button "23" .<br />
STRATO X (120V) 94<br />
(Rev. 1)