USER'S MANUAL<br />
General instructions for use<br />
4. Position the patient at temple support "4" (Figure 18), resting the<br />
chin on the support base 1. The hands should rest on the front<br />
handles. Ask the patient to bite the reference notch of bite "3" on rod<br />
"2" (Figure 18) with his incisors. In case of edentulous patients,<br />
he/her must rest the chin against the reference shoulder of the<br />
edentulous chin support.<br />
5. Instruct the patient to close the eyes.<br />
6. Press the ON button "20" of the laser centering device. Three<br />
laser beams will light up the mid-sagittal plane line "45", the<br />
horizontal line for the Frankfurt plane reference "46" (the plane that<br />
identifies a line that ideally links the ear hole - the auditory meatus -<br />
with the lower part of the orbital fossa Figure 18) and the canine<br />
plane line "47" (Figure 18). Position the patient's head in such a way<br />
as to ensure that the first two luminous beams fall in correspondence<br />
of the respective anatomical references.<br />
The luminous beam of the Frankfurt plane can be adjusted according<br />
to the height of the patient. This adjustment is made using knob "5"<br />
(Figure 18) located on the side of the tubehead.<br />
7. Also, close the temple support acting on the two knobs "6" (Figure<br />
18). The temple support will help the patient to stay in a correct<br />
position. Check that, during this phase, patient has not changed the<br />
position.<br />
8. Press key "23" to confirm. The confirmation request LED<br />
switches off. The luminous centering devices switch off and the<br />
rotating arm goes to its examination start position. Once alignment<br />
has been completed, the following message will be displayed:<br />
S T A R T E X A M<br />
x x k V x x m A 1 0 . 5 s<br />
x = value defined by the settings<br />
<br />
NOTE:<br />
The laser centering device remains lit until the laser centering device ON<br />
button "20"<br />
is pressed or, upon completion of the alignment<br />
procedure, key "23"<br />
. is pressed to start the exam procedure.<br />
9. Make the following recommendations to the patient: the mouth<br />
must remain closed, the tongue must be placed towards the<br />
palate, he/she must remain perfectly still and do not look at the<br />
rotating arm during the movements.<br />
(Rev. 1)<br />
71<br />
STRATO X (120V)