Strato X User Manual - Image Works
Strato X User Manual - Image Works
Strato X User Manual - Image Works
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
USER'S MANUAL<br />
Technical specifications<br />
Laser centering devices<br />
3 laser beams are used for the patient positioning; beams align mid Sagittal,<br />
Frankfurt and Canine Planes (please refer to relevant paragraphs for detailed<br />
explanation).<br />
Wave length<br />
Optical power of laser diode<br />
Optical power of the collimed beam<br />
Divergence<br />
Optical power on the working surface<br />
Laser class<br />
DNRO in 30 s application period<br />
Mechanical characteristics<br />
Focus-film distance (PAN, TMJ and<br />
Implant)<br />
Film size (PAN, TMJ and Implant)<br />
Focus film distance (CEPH)<br />
Film size (CEPH)<br />
Telescopic motorized column run<br />
Total height max.<br />
635 nm<br />
5 mW<br />
4.5 nW<br />
6.67 mRad<br />
< 3 mW<br />
3 A<br />
0.05 m<br />
20" (51 cm)<br />
15x30 cm flat cassette<br />
65" (165 cm)<br />
8"x10" (18x24 cm) and<br />
10"x12" (24x30 cm)<br />
26.4" (67 cm)<br />
91.3" (232 cm)<br />
Width x Length • 39.4" x 49.2" (100 x 125 cm) base<br />
machine<br />
• 69.7" x 49.2" (177 x 125 cm) with<br />
Ceph<br />
Weight • 135 kg base machine<br />
• 150 kg with Ceph<br />
Column weight<br />
Weight of arm support, rotating arm and<br />
tubehead<br />
Cassette holder weight<br />
Weight of the chin rest arm<br />
72 kg<br />
48 kg<br />
7 kg<br />
8 kg<br />
STRATO X (120V) 18<br />
(Rev. 1)