USER'S MANUAL<br />
Technical specifications<br />
<br />
WARNING:<br />
X-ray will be emitted during the execution of the following operations.<br />
Authorised technicians are therefore recommended to use the greatest<br />
caution and to comply with the safety regulations and laws of the of their<br />
country.<br />
7. Perform an exposure; the emitted parameters (kV and Time) can be<br />
read on the measuring instrument.<br />
<br />
NOTE:<br />
The performances are guaranteed if the measurement of kV and it is<br />
done with the invasive method.<br />
To quit this routine, press key "9"<br />
; the display will show:<br />
C L O S E C A S S E T T E<br />
T O P A N O R A M I C<br />
After closing the cassette holder to Panoramic position (the position is<br />
sensed by the unit through microswitch S24), the display will show the<br />
following message:<br />
P R E S S E N T E R<br />
Press key "23"<br />
; the unit will carry out the zeroing and will set<br />
itself in the patient entry position; the display will first show:<br />
P L E A S E W A I T . . .<br />
then<br />
x x k V x x m A 1 5 . 0 s<br />
and the unit returned at a standard mode.<br />
STRATO X (120V) 30<br />