USER'S MANUAL<br />

Technical specifications<br />

<br />

WARNING:<br />

X-ray will be emitted during the execution of the following operations.<br />

Authorised technicians are therefore recommended to use the greatest<br />

caution and to comply with the safety regulations and laws of the of their<br />

country.<br />

7. Perform an exposure; the emitted parameters (kV and Time) can be<br />

read on the measuring instrument.<br />

<br />

NOTE:<br />

The performances are guaranteed if the measurement of kV and it is<br />

done with the invasive method.<br />

To quit this routine, press key "9"<br />

; the display will show:<br />

C L O S E C A S S E T T E<br />

T O P A N O R A M I C<br />

After closing the cassette holder to Panoramic position (the position is<br />

sensed by the unit through microswitch S24), the display will show the<br />

following message:<br />

P R E S S E N T E R<br />

Press key "23"<br />

; the unit will carry out the zeroing and will set<br />

itself in the patient entry position; the display will first show:<br />

P L E A S E W A I T . . .<br />

then<br />

x x k V x x m A 1 5 . 0 s<br />

and the unit returned at a standard mode.<br />

STRATO X (120V) 30<br />

(Rev. 1)