USER'S MANUAL<br />

General instructions for use<br />

<br />

NOTE:<br />

If, during the exposure the patient moves or the unit collides with the<br />

patient or with another structure, or the operator realizes that the<br />

parameters set are incorrect, the X-ray button "30" must be released<br />

immediately, thus suspending emission of the X-rays and the movement<br />

of the arm.<br />

If this occurs, the following message will be displayed:<br />

E 5 6 1<br />

P R E S S R E S E T<br />

All the motors will be switched off and it will be possible, if necessary, to<br />

manually rotate the arm allowing the patient to move away from the unit.<br />

It is recommended that this movement be made with great care in<br />

order to prevent damage to the machine.<br />

Press button "9"<br />

. The following message will be displayed:<br />

M A C H I N E S E T T I N G<br />

Press key "23"<br />

. The following message will be displayed:<br />

W A I T F O R<br />

M A C H I N E S E T T I N G<br />

and then:<br />

R E P L A C E<br />

C A S S E T T E<br />

The system now returns to its initial position and the unit must be reset<br />

(cassette and parameters) and the patient repositioned.<br />

<br />

NOTE:<br />

Remove the cassette and change the film to prevent a double exposure of<br />

the film that would then provide non-diagnostic results.<br />

(Rev. 1)<br />