USER'S MANUAL<br />
General instructions for use<br />
<br />
WARNING:<br />
Clean the edentulous chin support, the handles and the temple support<br />
thoroughly after each examination.<br />
<br />
NOTE:<br />
If, during the exposure the patient moves or the unit collides with the<br />
patient or with another structure, or the operator realises that the<br />
parameters set are incorrect, the X-ray button "30" must be released<br />
immediately, thus suspending emission of the X-rays and the movement<br />
of the arm.<br />
If this occurs, the following message will be displayed:<br />
E 5 6 1<br />
P R E S S R E S E T<br />
All the motors will be switched off and it will be possible, if necessary, to<br />
manually rotate the arm and allow the patient to move away from the<br />
unit. It is recommended that this movement be made with great care<br />
in order to prevent damage to the machine.<br />
Press button "9"<br />
. The following message will be displayed:<br />
M A C H I N E S E T T I N G<br />
Press key "23"<br />
. The following message will be displayed:<br />
W A I T F O R<br />
M A C H I N E S E T T I N G<br />
and then:<br />
R E P L A C E<br />
C A S S E T T E<br />
The system now returns to its initial position and the unit must be reset<br />
(cassette and parameters) and the patient can be repositioned.<br />
<br />
NOTE:<br />
If the open mouth exposure is not completed, the closed mouth exposure<br />
must be repeated or the four complete pictures will not appear on the<br />
same plate.<br />
No warning message will be given if the cassette is removed but the film<br />
is not replaced, and another exposure is made. However, the film would<br />
provide non-diagnostic results due to the double exposure.<br />
STRATO X (120V) 64<br />
(Rev. 1)