USER'S MANUAL<br />
General instructions for use<br />
Reference lines<br />
45 Sagittal median line<br />
46 Frankfurt plane line<br />
Patient positioning and<br />
centering device<br />
1 Edentulous support<br />
4 Temple support rod<br />
5 Frankfurt plane laser<br />
adjustment knob<br />
6 Adjustment knobs for temple<br />
support<br />
Figure 16 – Open mouth exam positioning<br />
5. Instruct the patient to close the eyes.<br />
6. Press the laser centering device on key "20" . Three laser<br />
beams will light the sagittal median plane "45", the horizontal line for<br />
the Frankfurt plane reference "46" (Figure 16) and the canine plane<br />
line. Position the patient's head in such a way that the sagittal<br />
median plane is lit by the corresponding laser beam as per Figure 16.<br />
7. After having positioned the head, close the temple support acting on<br />
the two knobs "6" (Figure 16). The temple support will help the<br />
patient to stay in a correct position. Check that, during this phase,<br />
patient has not changed the position.<br />
<br />
NOTE:<br />
The laser centering device remains lit until the laser centering device ON<br />
button "20"<br />
is pressed or, upon completion of the alignment<br />
procedure, key "23"<br />
. is pressed to start the exam procedure.<br />
8. Advise the patient to remain perfectly still and do not look at<br />
the rotating arm during the movements.<br />
(Rev. 1)<br />
61<br />
STRATO X (120V)