USER'S MANUAL<br />
General instructions for use<br />
<br />
NOTE:<br />
The rotation of the arm and the emission of the X-rays will start with a<br />
delay of 2 seconds from pressing the X-ray button "30". Since the X-ray<br />
button is a "dead man" control, it must be kept pressed until the end<br />
of the exposure.<br />
The X-rays emission to the central part of the dental arch is suspended<br />
during the examination phase, the relative signals (sound and visual) are<br />
therefore also suspended.<br />
2. Once the exposure is completed the system will rotate back. When it<br />
stops, the following message will be displayed:<br />
P A T I E N T E X I T<br />
P R E S S → 0 ←<br />
Now remove the patient from the positioning device.<br />
3. Press button "10" . The machine will return to the patient<br />
entry position and the following message will be displayed<br />
P L E A S E W A I T . . .<br />
<br />
WARNING:<br />
During this phase the system stores into the internal memory some data.<br />
Switching off the unit during this phase can lead to the loss of some<br />
set-up data normally stored in the memory, which in turn may cause<br />
irreversible damages to the system itself.<br />
Do not switch off the unit in this phase.<br />
If required, switch off the unit only when the green LED "23"<br />
is lit.<br />
Then the following message will be displayed:<br />
R E P L A C E<br />
C A S S E T T E<br />
When the cassette has been removed, LED "23"<br />
and the following message will be displayed:<br />
will light up<br />
x x k V x x m A 1 0 . 5 s<br />
x = value defined by the settings<br />
That shows the values set for the exposure just made. A new<br />
exposure can now be made.<br />
(Rev. 1)<br />
63<br />
STRATO X (120V)