USER'S MANUAL<br />
General instructions for use<br />
5. Position the patient at temple support "4" (Figure 15). Rest the hands<br />
on the front handles.<br />
6. Instruct the patient to close the eyes.<br />
7. Press the laser centering device ON key "20" . Three laser<br />
beams will light the sagittal median plane "45", the horizontal line for<br />
the Frankfurt plane reference "46" (the plane that identifies a line<br />
that ideally links the ear hole - the auditory meatus - with the lower<br />
part of the orbital fossa in Figure 15) and the canine plane line.<br />
Position the patient's head in such a way that the first two light<br />
beams correspond with the relative anatomical parts.<br />
The luminous beam of the Frankfurt plane can be adjusted according<br />
to the height of the patient. This adjustment is made using knob "5"<br />
(Figure 15) located on the side of the tubehead.<br />
8. Also, close the temple support acting on the two knobs "6" (Figure<br />
15). The temple support will help the patient to stay in a correct<br />
position. Check that, during this phase, patient has not changed the<br />
position.<br />
9. Press key "23" to confirm. The confirmation request LED<br />
switches off. The luminous centering devices switch off and the<br />
rotating arm goes to its examination start position. Once alignment<br />
has been completed, the following message will be displayed:<br />
S T A R T E X A M<br />
x x k V x x m A 1 5 . 0 s<br />
x = value defined by the settings<br />
<br />
NOTE:<br />
The laser centering device remains lit until the laser centering device ON<br />
button "20"<br />
is pressed or, upon completion of the alignment<br />
procedure, key "23"<br />
. is pressed to start the exam procedure.<br />
10. Make the following recommendations to the patient: the lips<br />
must remain closed, he/she must remain perfectly still and do<br />
not look at the rotating arm during the movements.<br />
(Rev. 1)<br />
57<br />
STRATO X (120V)