USER'S MANUAL<br />
Technical specifications<br />
2. Press keys "10" , "28" and "24" at the same<br />
time. LED’s of "patient type", "patient size" and "Arch" switch<br />
off and the display shows the following two messages alternatively:<br />
and<br />
R E M O V E<br />
C H I N R E S T<br />
C L O S E<br />
T E M P L E S U P P O R T<br />
<br />
NOTE:<br />
The following operations is a confirmation that the above points have<br />
been performed.<br />
3. Carry out the actions mentioned above; press key "23" ;<br />
the unit will carry out a movement to reach the "zero" position.<br />
During this phase, the display shows:<br />
P L E A S E W A I T . . .<br />
Once the "zero" position is reached, the display will show the<br />
following message:<br />
O P E N<br />
C A S S E T T E U N I T<br />
4. After opening the cassette holder to Ceph position (this action is<br />
sensed by the unit which monitors the status of microswitches S25),<br />
the display shows:<br />
P L E A S E W A I T . . .<br />
In the mean time, the unit sets the primary collimator to position<br />
8"x10" symmetric (Slot #5) and the soft tissue filter in position "not in<br />
field"; at the end of the positioning phase, the display shows:<br />
R X P A R A M E T E R S<br />
7 0 k V 0 8 m A 1 . 0 0 S<br />
STRATO X (120V) 28<br />