USER'S MANUAL<br />
General instructions for use<br />
6.6.1 Getting started<br />
Press key "19<br />
to select the Biaxial TMJ examination.<br />
The following configuration is automatically set:<br />
• ADULT : the upper led of key "11" lights up<br />
• MEDIUM SIZE: the central led of key "12" lights up<br />
• MEDIUM ARCH: the central led of key "13" lights up<br />
• Biaxial TMJ: the LED of key "19" lights up<br />
LED "23" lights up and the display shows the default exposure<br />
parameters if this is the first Biaxial TMJ exposure made, or the exposure<br />
parameters (kV and mA) used in the last exposure. For example:<br />
7 4 k V 1 0 m A 1 0 . 5 s<br />
Once the machine has been set, place the standard chin support in<br />
position if it had been previously removed (see notes in paragraph 6.3).<br />
<br />
NOTE:<br />
STRATO X is based on a dentition and ascending rami shape as defined<br />
by U. Welander et al., Dentomaxillofacial Radiology, 1989, Vol. 18, May.<br />
This paper, based on statistic study, enstabilishes a form for the the<br />
dentomaxillofacial complex that it is assumed as "standard". A projection<br />
geometry that maintains a constant magnification of 1.23 throughout the<br />
exposure of this shape has been applied to the STRATO X. Patient’s<br />
anatomy can differ in a significant way from the statistical model, so the<br />
magnification factor is not maintained and can be different from that<br />
value. Based on his experience and competence, the user has to judge<br />
this variation.<br />
IN ANY CASE, THE TMJ RADIOGRAPHY CANNOT BE USED TO<br />
PERFORM CALCULATIONS OF DISTANCES, ANGLES ETC. ON THE<br />
FILM.<br />
STRATO X (120V) 66<br />
(Rev. 1)