Process Unit 73 LF - Knick
Process Unit 73 LF - Knick
Process Unit 73 LF - Knick
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong><br />
Probe Rinsing<br />
You can only make use of the probe rinsing function<br />
if your <strong>Process</strong> <strong>Unit</strong> is equipped with option<br />
352. Without this option, the menu reads ”Probe<br />
Rinsing (Optional)”. This menu item cannot be selected.<br />
Probe rinsing is used, for example, for automatic<br />
rinsing and cleaning of the conductivity cell.<br />
To do so, a rinsing cycle is started.<br />
A rinsing cycle can be started:<br />
<br />
<br />
<br />
<br />
timer controlled after expiration of the user defined<br />
interval,<br />
manually in the menu,<br />
by a current pulse (see page 9–46) at the current<br />
input (if current input has been set as control<br />
input, see page 9–43),<br />
remote-controlled via interface (see page<br />
9–49).<br />
You can define an interval in the range 0.1 to<br />
999.9 h. The interval extends from the start of one<br />
rinsing cycle to the start of the next rinsing cycle.<br />
To turn off the automatic rinsing cycle, set the interval<br />
to ”000.0 h”.<br />
Before starting a rinsing cycle, you must enter the<br />
individual step durations on the Parameter Setting<br />
level, menu item ”Probe Rinsing” (see page 9–47)!<br />
A rinsing cycle comprises the following steps:<br />
<br />
<br />
Start of Rinsing Cycle:<br />
NAMUR contact ”functional check” and contact<br />
”probe” are activated, output current 1 (and 2)<br />
and controller output are frozen, limits are disabled,<br />
and menus are locked, the<br />
interval timer is reset.<br />
Rinsing Lead Time:<br />
User defined waiting time until contact ”rinsing”<br />
closes. This allows you to monitor response<br />
times of the valve ”probe”, for example .<br />
<br />
First Rinsing:<br />
The contact ”rinsing” is closed for the (user defined)<br />
duration of first rinsing.<br />
9–44 Capabilities