Probe Rinsing<br />

You can only make use of the probe rinsing function<br />

if your <strong>Process</strong> <strong>Unit</strong> is equipped with option<br />

352. Without this option, the menu reads ”Probe<br />

Rinsing (Optional)”. This menu item cannot be selected.<br />

Probe rinsing is used, for example, for automatic<br />

rinsing and cleaning of the conductivity cell.<br />

To do so, a rinsing cycle is started.<br />

A rinsing cycle can be started:<br />

<br />

<br />

<br />

<br />

timer controlled after expiration of the user defined<br />

interval,<br />

manually in the menu,<br />

by a current pulse (see page 9–46) at the current<br />

input (if current input has been set as control<br />

input, see page 9–43),<br />

remote-controlled via interface (see page<br />

9–49).<br />

You can define an interval in the range 0.1 to<br />

999.9 h. The interval extends from the start of one<br />

rinsing cycle to the start of the next rinsing cycle.<br />

To turn off the automatic rinsing cycle, set the interval<br />

to ”000.0 h”.<br />

Before starting a rinsing cycle, you must enter the<br />

individual step durations on the Parameter Setting<br />

level, menu item ”Probe Rinsing” (see page 9–47)!<br />

A rinsing cycle comprises the following steps:<br />

<br />

<br />

Start of Rinsing Cycle:<br />

NAMUR contact ”functional check” and contact<br />

”probe” are activated, output current 1 (and 2)<br />

and controller output are frozen, limits are disabled,<br />

and menus are locked, the<br />

interval timer is reset.<br />

Rinsing Lead Time:<br />

User defined waiting time until contact ”rinsing”<br />

closes. This allows you to monitor response<br />

times of the valve ”probe”, for example .<br />

<br />

First Rinsing:<br />

The contact ”rinsing” is closed for the (user defined)<br />

duration of first rinsing.<br />

9–44 Capabilities