Process Unit 73 LF - Knick
Process Unit 73 LF - Knick
Process Unit 73 LF - Knick
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong><br />
When the cursor is on the first digit, increases<br />
the input range by one decade.<br />
Input range changes to 00.00 – 99.99 mS/cm.<br />
The sequence ”104” remains on the display,<br />
shifted by one decimal position. The last digit (”7”)<br />
is erased.<br />
When the cursor is on the last digit, decreases<br />
the input range by one decade.<br />
Input range changes to 000.0 – 999.9 S/cm.<br />
The sequence ”047” remains on the display,<br />
shifted by one decimal position. The first digit (”1”)<br />
is erased.<br />
The digits that are shifted out of the display are<br />
erased (set to zero). That is, only the displayed<br />
4-digit value is edited.<br />
Pressing restores the old value if you have<br />
not yet pressed .<br />
Press the scrolling keys or to scroll<br />
from 0 through 9 and change the sign.<br />
How to change a sign<br />
If the entry value has a sign, the flashing cursor<br />
can be placed on it using .<br />
Pressing or switches between ”+” and<br />
”–”.<br />
An Example<br />
In this example we want to change the temperature<br />
alarm ”Warning Limit Hi” from 50 to 67 C.<br />
Press twice until the flashing cursor is on digit<br />
”5”.<br />
Press once (”6”).<br />
Press once: The flashing cursor is on digit ”0”.<br />
Press three times (”7”).<br />
2–6 Operation