NAMUR Contacts<br />
The three NAMUR contacts functional check,<br />
warning (maintenance required) and failure are<br />
provided as standard.<br />
<br />
<br />
<br />
Functional check is activated:<br />
during calibration (),<br />
during maintenance (): current source,<br />
meas. point maintenance,<br />
during parameter setting () on the<br />
Operator level (opl) and the Administrator level<br />
(adm),<br />
or during an automatic rinsing cycle.<br />
Warning (maintenance required) is activated<br />
if a value has exceeded (or fallen below, resp.)<br />
a preset ”Warning Limit Hi” or ”Warning Limit<br />
Lo”, or if other warning messages have been<br />
activated.<br />
That means that the measuring installation is<br />
still functioning properly but should be maintained,<br />
or that certain process parameters have<br />
reached a value that requires intervention.<br />
Warning is disabled during ”Functional Check”.<br />
Failure is activated<br />
if a value has exceeded (or fallen below, resp.)<br />
a preset ”Failure Limit Hi” or ”Failure Limit Lo”,<br />
if the range limits of the <strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong><br />
have been exceeded, or if other failure messages<br />
have been activated.<br />
That means that the measuring installation is<br />
not properly functioning any more, or that certain<br />
process parameters have reached a critical<br />
value.<br />
Failure is disabled during ”Functional Check”.<br />
You can set the three NAMUR contacts as normally<br />
open contacts (active: closed) or as normally<br />
closed contacts (active: open).<br />
For safe operation, the NAMUR contacts must be<br />
set as normally closed contacts.<br />
Only this ensures that power failure will activate<br />
an alarm message!<br />
You can define a delay time for both the warning<br />
and the failure contact . If an alarm message is<br />
released, the contact will only be activated after<br />
this preset delay time.<br />
As delivered, the relay contacts are suitable for<br />
low signal currents (down to approx. 1 mA).<br />
If currents above approx. 100 mA are switched,<br />
the gold plating is destroyed during the switching<br />
process. After that, the contacts will not reliably<br />
switch low currents.<br />
Capabilities 9–29