Process Unit 73 LF - Knick
Process Unit 73 LF - Knick
Process Unit 73 LF - Knick
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
14 Specifications<br />
Conductivity input<br />
either 4-electrode or 2-electrode cells<br />
Ranges<br />
Conductivity<br />
0.001 S/cm to 2,000 mS/cm<br />
Resistivity (1/)<br />
0.5 cm to 1,000 Mcm<br />
Concentration<br />
0.00 to 200.0 % by wt<br />
Display range<br />
0.1 S ⋅ c to 1,999 mS ⋅c (cell constant c = 0.0050 to 200.0 cm -1 )<br />
Accuracy ( 1 count)<br />
< 1 % of measured value + 0.2 S ⋅ c<br />
max. permissible direct voltage against signal ground 2 V<br />
(especially when using stray-field probes, equipotential bonding is<br />
required)<br />
Resolution<br />
0.000 S/cm c = 0.0050 to 0.1199 cm -1<br />
0.00 S/cm c = 0.1200 to 1.199 cm -1<br />
0.0 S/cm c = 1.200 to 11.99 cm -1<br />
0.000 mS/cm c = 12.00 to 119.9 cm -1<br />
0.00 mS/cm c = 120.0 to 200.0 cm -1<br />
Cell standardization<br />
Operating modes*)<br />
automatic, by cell constant determination<br />
with NaCl or KCl solution<br />
entry of individual conductivity values<br />
for cell constant determination<br />
direct entry of cell constant<br />
calibration by sampling<br />
Permissible cell constant 0.0050 to 200.0 cm -1<br />
Concentration determination<br />
(Option 359, 360, 382)<br />
Calculation and display of concentration [% by wt.] from the<br />
conductivity and temperature values of specified substance<br />
solutions (see tables in annex)<br />
customer-specific tables on request (opt. 360)<br />
Temperature input<br />
Pt 100 / Pt 1000 / Ni 100<br />
2 or 3-wire connection<br />
Range<br />
Pt 100/1000<br />
-50.0 to +250.0 C<br />
Ni 100<br />
-50.0 to +180.0 C<br />
Accuracy ( 1 count)<br />
Temperature compensation *)<br />
media-dependent<br />
Current input<br />
Accuracy<br />
< 0.2 % of measured value, 0.2 K<br />
automatic with Pt 100/Pt 1000/Ni 100<br />
manual -50.0 to +250.0 C<br />
Operating modes:<br />
without<br />
linear 0.00 to 20.00 %/K, reference temperature user-defined<br />
natural waters to EN 27888<br />
optional: to customer specification (option 361)<br />
ultrapure water with traces of impurity (option 392)<br />
0(4) to 20 mA<br />
evaluation 0 to 100 %<br />
input resistance 50 , overload 100 mA<br />
e.g. for limit monitoring<br />
In conjunction with power output complete 2-wire loop,<br />
e.g. for flow or level meter<br />
< 1 % full scale<br />
Specifications 14–1