14 Specifications<br />

Conductivity input<br />

either 4-electrode or 2-electrode cells<br />

Ranges<br />

Conductivity<br />

0.001 S/cm to 2,000 mS/cm<br />

Resistivity (1/)<br />

0.5 cm to 1,000 Mcm<br />

Concentration<br />

0.00 to 200.0 % by wt<br />

Display range<br />

0.1 S ⋅ c to 1,999 mS ⋅c (cell constant c = 0.0050 to 200.0 cm -1 )<br />

Accuracy ( 1 count)<br />

< 1 % of measured value + 0.2 S ⋅ c<br />

max. permissible direct voltage against signal ground 2 V<br />

(especially when using stray-field probes, equipotential bonding is<br />

required)<br />

Resolution<br />

0.000 S/cm c = 0.0050 to 0.1199 cm -1<br />

0.00 S/cm c = 0.1200 to 1.199 cm -1<br />

0.0 S/cm c = 1.200 to 11.99 cm -1<br />

0.000 mS/cm c = 12.00 to 119.9 cm -1<br />

0.00 mS/cm c = 120.0 to 200.0 cm -1<br />

Cell standardization<br />

Operating modes*)<br />

automatic, by cell constant determination<br />

with NaCl or KCl solution<br />

entry of individual conductivity values<br />

for cell constant determination<br />

direct entry of cell constant<br />

calibration by sampling<br />

Permissible cell constant 0.0050 to 200.0 cm -1<br />

Concentration determination<br />

(Option 359, 360, 382)<br />

Calculation and display of concentration [% by wt.] from the<br />

conductivity and temperature values of specified substance<br />

solutions (see tables in annex)<br />

customer-specific tables on request (opt. 360)<br />

Temperature input<br />

Pt 100 / Pt 1000 / Ni 100<br />

2 or 3-wire connection<br />

Range<br />

Pt 100/1000<br />

-50.0 to +250.0 C<br />

Ni 100<br />

-50.0 to +180.0 C<br />

Accuracy ( 1 count)<br />

Temperature compensation *)<br />

media-dependent<br />

Current input<br />

Accuracy<br />

< 0.2 % of measured value, 0.2 K<br />

automatic with Pt 100/Pt 1000/Ni 100<br />

manual -50.0 to +250.0 C<br />

Operating modes:<br />

without<br />

linear 0.00 to 20.00 %/K, reference temperature user-defined<br />

natural waters to EN 27888<br />

optional: to customer specification (option 361)<br />

ultrapure water with traces of impurity (option 392)<br />

0(4) to 20 mA<br />

evaluation 0 to 100 %<br />

input resistance 50 , overload 100 mA<br />

e.g. for limit monitoring<br />

In conjunction with power output complete 2-wire loop,<br />

e.g. for flow or level meter<br />

< 1 % full scale<br />

Specifications 14–1