PARAMETER Commands: Query settings and set parameters<br />
With the parameter commands, all functions of the <strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong> can be defined via the computer<br />
interface (with the exception of the interface transmission parameters).<br />
With the parameter commands, all instrument parameters can be read and written!<br />
Therefore, the correctness of the transmitted commands is particularly important.<br />
Transmission in the point-to-point mode is not secured with checksums. To avoid incorrect<br />
settings, it is therefore advisable to read back important parameters for a comparison.<br />
With the first write command, the controlling computer (PC, SPS, ...) assumes control<br />
of the <strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong>. Many security prompts must then be implemented in the<br />
computer! With the “WCOMIN0” command (go to local) the computer returns control<br />
to the <strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong>. The <strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong> restarts in the measuring mode.<br />
Read commands cause no status changes and do not influence the system functions.<br />
The <strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong> continues to be in control.<br />
When the write protection is switched on, any write attempts without switching off the<br />
write protection beforehand or with an invalid passcode are entered in the logbook.<br />
As delivered, write protection is switched off.<br />
If parameter setting is called up with the device keypad, the NAMUR message Functional check is<br />
set. Warning and failure contacts are deactivated until parameter setting is completed.<br />
If device parameters are changed via the RS 485 interface, all messages are released. As a result,<br />
temporary messages may occur when changing parameters which would have been suppressed had<br />
the keypad been used.<br />
WCOM01<br />
WCOU1<br />
With the “WCOM01” interface command, the device can be switched to the parameter<br />
setting mode. The NAMUR message Functional check is then set and with it the warning<br />
and failure contact also deactivated in the interface mode.<br />
Return to the measuring mode with “WCOM00”.<br />
If all device functions are to be frozen during parameter setting, use the “WCOU1”<br />
command. The Functional Check is set, and the warning and failure contact deactivated.<br />
The output current and controller are also frozen and the limit-value contacts<br />
are inactive.<br />
Unfreeze the device functions with “WCOU0”.<br />
Point of Measurement<br />
RPUAM<br />
WPUAM0<br />
WPUAM1<br />
RPUAW<br />
WPUAWaaaaaaaaaaaaaaa<br />
Query marker<br />
Set marker ”Off”<br />
Set marker ”On”<br />
Query entered point of measurement<br />
Enter point of measurement<br />
a = ASCII character: blank, ”0” ... ”9”, ”A ” ... ”Z”, ”–”, ”+”, ”/”<br />
Interface Commands 12–9