23.03.2014 Views

Process Unit 73 LF - Knick

Process Unit 73 LF - Knick

Process Unit 73 LF - Knick

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong><br />

STATUS Commands: Query messages and states<br />

With the status commands, the device messages, such as the NAMUR messages Functional check,<br />

Warning (maintenance required) and Failure can be read out, device states monitored and the records<br />

queried. With the status commands, data can be accessed which can be used for QM documentation<br />

to ISO 9000. Status commands are read commands. As a result, the device status of the<br />

<strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong> is not changed.<br />

Command Function Response<br />

Meaning<br />

RSF1<br />

RSFA<br />

RSW1<br />

RSWA<br />

Query first failure message<br />

Query all failure messages<br />

Query first warning message<br />

Query all warning messages<br />

RSP Query device status (”menu”) 00 measuring mode<br />

01 parameter setting opl, adm<br />

02 calibration cal<br />

08 maintenance maint<br />

10 measuring mode, probe rinsing<br />

running, started by timer<br />

11 parameter setting opl, adm &<br />

probe rinsing running, started<br />

by timer<br />

18 maintenance, probe rinsing<br />

running, started manually<br />

RSL Query limit messages 0 no limit message<br />

1 limit 1 enabled<br />

2 limit 2 enabled<br />

3 both limits enabled<br />

12–6 Interface Commands