<strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong><br />
Controller Function<br />
You can only make use of the controller function if<br />
your unit is equipped with option 353 for the digital<br />
controller or options 350 and 483 for the analog<br />
controller (output 2).<br />
Without this option, the Parameter Setting menu<br />
will only read “Limits” instead of “Limits / Controller”<br />
and “Output Current 2” or “Output Current 2<br />
(Optional)” instead of “Output 2 / Controller” for the<br />
analog controller.<br />
Digital Controller<br />
The parameters of the digital controller are set<br />
with the “Limit Values / Controller” menu item.<br />
The bidirectional PI controller enables quasi-continuous<br />
(switched) control.<br />
Two controller types can be selected:<br />
Type A: Pulse length controller (see Pg. 9–37)<br />
<br />
Type B: Pulse frequency controller (see<br />
Pg. 9–37)<br />
Analog Controller<br />
The parameters of the analog controller are set<br />
with the “Output 2 / Controller” menu item.<br />
Three controller types can be selected:<br />
Type A: 3-way mixing valve (see Pg. 9–38)<br />
<br />
<br />
Type B: Straightway valve (< setpoint) (see<br />
Pg. 9–39)<br />
Type C: Straightway valve ( > setpoint) (see<br />
Pg. 9–40)<br />
Controlled Variables<br />
You can define as controlled variables:<br />
<br />
<br />
<br />
conductivity value<br />
resistivity<br />
measured temperature value (C)<br />
Since the relationship between conductivity and<br />
concentration is strongly nonlinear and partially<br />
ambiguous, concentration cannot be selected as<br />
controlled variable.<br />
9–32 Capabilities