Remote Interface Operation<br />

You can only make use of the interface if your <strong>Process</strong><br />

<strong>Unit</strong> is equipped with option 351.<br />

Without this option, the menu reads ”Interface<br />

(Optional)”. This menu item cannot be selected.<br />

When the RS 485 interface is used, terminal 15<br />

(RS 485 shield) must be grounded to meet the radio<br />

interference limits (according to German decree<br />

243/91). Do not use the protective conductor<br />

for grounding!<br />

To operate the <strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong> at a PC, a commercial<br />

RS 232 C/RS 485 interface converter can<br />

be used.<br />

What you can do with the interface<br />

The serial RS 485 interface allows you to:<br />

<br />

<br />

<br />

<br />

read out all measured values<br />

query instrument status including limit and<br />

alarm messages, instrument diagnostics and<br />

logbook<br />

perform complete parameter setting<br />

start a rinsing cycle via remote control<br />

For description of the complete command set and<br />

the transmission protocol, refer to chapter 12.<br />

If the instrument is in remote status during interface<br />

operation, the reading ”Remote” appears in<br />

the upper right corner of the display in measuring<br />

mode.<br />

The keypad is locked for all entries!<br />

When the instrument is in measuring mode, you<br />

can press to return to ”local” status. (You<br />

are prompted for confirmation.) The keypad will be<br />

enabled.<br />

The interface can be defined for:<br />

<br />

<br />

point-to-point operation<br />

(<strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong> connected with a controller,<br />

e.g. PC) or<br />

bus operation with up to 31 instruments and a<br />

controller (e.g. PC) at a bus.<br />

Capabilities 9–49