Remote Interface Operation<br />
You can only make use of the interface if your <strong>Process</strong><br />
<strong>Unit</strong> is equipped with option 351.<br />
Without this option, the menu reads ”Interface<br />
(Optional)”. This menu item cannot be selected.<br />
When the RS 485 interface is used, terminal 15<br />
(RS 485 shield) must be grounded to meet the radio<br />
interference limits (according to German decree<br />
243/91). Do not use the protective conductor<br />
for grounding!<br />
To operate the <strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong> at a PC, a commercial<br />
RS 232 C/RS 485 interface converter can<br />
be used.<br />
What you can do with the interface<br />
The serial RS 485 interface allows you to:<br />
<br />
<br />
<br />
<br />
read out all measured values<br />
query instrument status including limit and<br />
alarm messages, instrument diagnostics and<br />
logbook<br />
perform complete parameter setting<br />
start a rinsing cycle via remote control<br />
For description of the complete command set and<br />
the transmission protocol, refer to chapter 12.<br />
If the instrument is in remote status during interface<br />
operation, the reading ”Remote” appears in<br />
the upper right corner of the display in measuring<br />
mode.<br />
The keypad is locked for all entries!<br />
When the instrument is in measuring mode, you<br />
can press to return to ”local” status. (You<br />
are prompted for confirmation.) The keypad will be<br />
enabled.<br />
The interface can be defined for:<br />
<br />
<br />
point-to-point operation<br />
(<strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong> connected with a controller,<br />
e.g. PC) or<br />
bus operation with up to 31 instruments and a<br />
controller (e.g. PC) at a bus.<br />
Capabilities 9–49