Automatic Calibration with Standard<br />
Calibration Solution<br />
For automatic calibration, the cell is immersed in a<br />
standard calibration solution.<br />
From the measured conductance and temperature,<br />
the <strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong> automatically calculates<br />
the cell constant.<br />
Temperature dependence of the calibration solution’s<br />
conductance is taken into account by the<br />
<strong>Process</strong> <strong>Unit</strong>.<br />
During calibration, output current (1 and 2) and<br />
controller output are frozen at their last values;<br />
limit contacts are disabled, contact ”probe” is enabled.<br />
What you have to know for calibration<br />
Use only fresh calibration solutions!<br />
The calibration solution must be selected during<br />
parameter setting (see page 9–13).<br />
Accuracy of calibration depends on exact detection<br />
of the calibration solution’s temperature: Using<br />
the measured or entered temperature, the <strong>Process</strong><br />
<strong>Unit</strong> <strong>73</strong> <strong>LF</strong> determines the setpoint for the calibration<br />
solution from a stored chart.<br />
Note the response time of the temperature probe!<br />
For exact determination of the cell constant, wait<br />
until the temperatures of temperature probe and<br />
calibration solution have balanced.<br />