Automatic Calibration with Standard<br />

Calibration Solution<br />

For automatic calibration, the cell is immersed in a<br />

standard calibration solution.<br />

From the measured conductance and temperature,<br />

the <strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong> automatically calculates<br />

the cell constant.<br />

Temperature dependence of the calibration solution’s<br />

conductance is taken into account by the<br />

<strong>Process</strong> <strong>Unit</strong>.<br />

During calibration, output current (1 and 2) and<br />

controller output are frozen at their last values;<br />

limit contacts are disabled, contact ”probe” is enabled.<br />

What you have to know for calibration<br />

Use only fresh calibration solutions!<br />

The calibration solution must be selected during<br />

parameter setting (see page 9–13).<br />

Accuracy of calibration depends on exact detection<br />

of the calibration solution’s temperature: Using<br />

the measured or entered temperature, the <strong>Process</strong><br />

<strong>Unit</strong> <strong>73</strong> <strong>LF</strong> determines the setpoint for the calibration<br />

solution from a stored chart.<br />

Note the response time of the temperature probe!<br />

For exact determination of the cell constant, wait<br />

until the temperatures of temperature probe and<br />

calibration solution have balanced.<br />

Calibration 6–5