4 Maintenance Menu<br />

What you can do in the Maintenance<br />

menu<br />

The Maintenance menu provides all functions for<br />

sensor maintenance and adjustment of connected<br />

units.<br />

Access to the Maintenance menu can be protected<br />

by a passcode.<br />

<br />

<br />

<br />

<br />

<br />

<br />

Measurement point maintenance allows to dismount<br />

the cell.<br />

The probe rinsing function (option 352) permits<br />

automatic rinsing and cleaning of the cell: see<br />

page 9–44.<br />

Resistance measurement permits direct display<br />

of the ohmic resistance at the measuring input.<br />

The current source allows to manually adjust<br />

the output currents (1 and 2) for adjusting and<br />

checking connected peripheral devices (such<br />

as indicator or recorder).<br />

Temperature probe adjustment allows individual<br />

calibration of the temperature probe to increase<br />

accuracy of conductivity measurement (only<br />

effective with process medium TC enabled).<br />

If the unit is equipped with a controller function<br />

(option 353), controller output Y (manipulated<br />

variable) can be entered manually.<br />

Only with option 352: In the ”Measurement Point<br />

Maintenance” submenu, the ”probe” contact is active.<br />

A timer controlled rinsing cycle will not be<br />

started (see page 9–44).<br />

Maintenance 4–1