23.03.2014 Views

Process Unit 73 LF - Knick

Process Unit 73 LF - Knick

Process Unit 73 LF - Knick

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong><br />

Error Message<br />

(Display in Diagnostics Menu<br />

“Message List”)<br />

Warn Cal Temp<br />

Warn Cell Const<br />

Possible Error Causes and Remedies<br />

Calibration temperature out of range<br />

Cell constant < 0.005 or > 200 1/cm during calibration<br />

Warn Control Parameters Parameter error for controller (see page 9–41)<br />

Warn Current1 20 mA<br />

Warn Current1 Span<br />

Warn Current2 20 mA<br />

Warn Current2 Span<br />

Current output 1: Output current below defined initial value<br />

Current output 1: Output current above defined end value<br />

Current output 1: Initial and end value too close<br />

Current output 2: Output current below defined initial value<br />

Current output 2: Output current above defined end value<br />

Current output 2: Initial and end value too close<br />

Warn Current Par Parameter error for current output (see page 9–22)<br />

Warn Hi Cell Const<br />

Warn Hi Conc Value<br />

Warn Hi Conduct Value<br />

Warn Hi Current Inp<br />

Warn Hi Feed Time<br />

Warn Hi Temp<br />

Warn Interface<br />

Warn Lo Cell Const<br />

Warn Lo Conc Value<br />

Warn Lo Conduct Value<br />

Warn Lo Current Inp<br />

Warn Lo Temp<br />

Warn Ref Temp<br />

Cell constant above warning limit<br />

Concentration above warning limit<br />

Conductivity above warning limit<br />

Input current above warning limit<br />

Controller: Feed time above warning limit<br />

Measured temperature above warning limit<br />

Interface error: Parity or framing error<br />

Cell constant below warning limit<br />

Concentration below warning limit<br />

Conductivity below warning limit<br />

Input current below warning limit<br />

Measured temperature below warning limit<br />

Reference temperature < –50 C or > 250 C<br />

Warn RS485 Bus Address Interface error: Invalid device address entered (0 or >31)<br />

Warn RS485 Overflow<br />

Warn RS485 Parameter<br />

Warn RS485 Syntax<br />

Warn Sensor Unstable<br />

Warn TC<br />

Interface error: Buffer overflow, too many characters received<br />

without message terminator<br />

Interface error: Command parameter error<br />

Interface error: Command Syntax error or<br />

command not available<br />

Stable end value not obtained during calibration<br />

Temperature coefficient < 0 or > 20 %/K<br />

11–2 Error Messages