Process Unit 73 LF - Knick
Process Unit 73 LF - Knick
Process Unit 73 LF - Knick
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong><br />
Parameter Setting menu<br />
passcode protection<br />
point of measurement<br />
reference temperature<br />
rinsing cycle<br />
rinsing lead time<br />
scrolling key<br />
secondary display<br />
second rinsing<br />
temperature coefficient<br />
temperature compensation<br />
Viewing level<br />
wait position<br />
warning (maintenance required)<br />
The Parameter Setting menu is divided into three<br />
submenus: Viewing level (view), Operator level<br />
(opl) and Administrator level (adm).<br />
The passcode protection protects access to calibration,<br />
maintenance, Operator and Administrator<br />
level. The passcodes can be edited or disabled on<br />
the Administrator level.<br />
Can be defined for identifying the <strong>Process</strong> <strong>Unit</strong>;<br />
can be displayed in the diag menu or read out via<br />
remote interface<br />
With temperature compensation enabled, the<br />
measured value is calculated for the reference<br />
temperature (usually 20 or 25 C) using the temperature<br />
coefficient.<br />
User defined sequence for cleaning the cell or<br />
other sensors; controls contacts ”probe”, ”rinsing”<br />
and ”cleaning”<br />
User defined time at the beginning of the rinsing<br />
cycle, after closing of contact ”probe”, before closing<br />
of contact ”rinsing”<br />
and keys for selecting menu lines or entering<br />
numeric digits<br />
Two small displays that appear at the lower left<br />
and right sides of the main display during measuring<br />
mode. The process variables displayed can be<br />
selected using / and / .<br />
User defined time during which contact ”rinsing” is<br />
closed at the end of the rinsing cycle<br />
With temperature compensation enabled, the<br />
measured value is calculated for the reference<br />
temperature using the temperature coefficient.<br />
Calculates the measured conductivity value for a<br />
reference temperature.<br />
”view” – menu level of parameter setting; display<br />
of all configuration settings, no editing possible<br />
Position between ”Second Rinsing” and ”Measurement<br />
Lead Time” where the probe remains as long<br />
as a start current of 10 to 20 mA is applied to the<br />
current input (only when current input has been<br />
set as control input)<br />
Alarm message and NAMUR contact;<br />
means that measuring equipment still operates<br />
properly but should be maintained, or that certain<br />
process parameters have reached a value that<br />
requires intervention.<br />
Warning is not enabled during ”functional check”.<br />
17–4 Technical Terms