<strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong><br />

Cells<br />

Model SE 600 (4-electrode cell) by Kemotron<br />

Cell constant<br />

typically 0.14 to 0,.8 cm –1 (exact value printed on rating plate)<br />

Range 0.5 S to 600 mS/cm 1)<br />

Material Body PTFE, stainless steel AISI 316 (≈ 1.4401)<br />

Electrodes stainless steel AISI 316 (≈ 1.4401)<br />

Max. temperature<br />

Max. pressure<br />

210 C<br />

50 bars<br />

Temperature probe Pt 1000 (IEC Class A)<br />

Cable Length 5 m<br />

Connection<br />

wire end ferrules<br />

Dimensions see dimension drawing Fig. 14–1<br />

Model SE 602*) (4-electrode cell) by Kemotron<br />

Cell constant<br />

typically 0.14 to 0.38 cm –1 (exact value printed on rating plate)<br />

Range 0.5 S to 600 mS/cm 1)<br />

Material Body PTFE<br />

Electrodes<br />

tantalum<br />

Max. temperature<br />

Max. pressure<br />

120 C<br />

20 bars<br />

Temperature probe Pt 1000 (IEC Class A)<br />

Cable Length 5 m<br />

Connection<br />

wire end ferrules<br />

Dimensions see dimension drawing Fig. 14–2<br />

*) For measurements in ungrounded media, equipotential bonding required<br />

1) up to 2,000 mS, however with reduced accuracy<br />

14–4 Specifications