23.03.2014 Views

Process Unit 73 LF - Knick

Process Unit 73 LF - Knick

Process Unit 73 LF - Knick

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Process</strong> <strong>Unit</strong> <strong>73</strong> <strong>LF</strong><br />

Model ZU 0071*) (2-electrode cell) by Endress + Hauser<br />

Cell constant approx. 1 cm –1<br />

Range<br />

approx. 5 S/cm to 15 mS/cm<br />

Material Body PES<br />

Electrodes<br />

special graphite, titanium<br />

Max. temperature<br />

Max. pressure<br />

Temperature probe<br />

150 C<br />

16 bars<br />

Pt 100 (in titanium sleeve)<br />

Cable Length 5 m<br />

Connection<br />

wire end ferrules<br />

Dimensions see dimension drawing Fig. 14–6<br />

*) For measurements in ungrounded media, equipotential bonding required<br />

14–6 Specifications