Cleaning:<br />
The contact ”cleaning” is closed for the (user<br />
defined) duration of cleaning.<br />
<br />
Second Rinsing:<br />
The contact ”rinsing” is closed for the (user defined)<br />
duration of second rinsing.<br />
<br />
Wait Position:<br />
If current input has been set as control input,<br />
the probe remains in wait position as long as<br />
the start current of 10 to 20 mA is applied to the<br />
current input.<br />
Wait position can only be controlled via the current<br />
input. If current input has been set as signal input,<br />
wait position is omitted.<br />
<br />
Measurement Lead Time:<br />
The contact ”probe” is deactivated.<br />
Now, the user defined waiting time until rinsing<br />
cycle is terminated starts running.<br />
After expiration, ”functional check” will be<br />
turned off.<br />
If you have set one of these steps to 0000 s, it is<br />
omitted completely.<br />
Pressing during the rinsing cycle displays<br />
the measured value for approx. 5 s.<br />