118. Lighting and Appliance Testing Laboratory<br />

Calibration / Instrumentation & Control Section<br />

Department <strong>of</strong> Energy<br />

Address : Commonwealth Avenue, Diliman<br />

Quezon City<br />

Telephone No. : (632) 927-7137<br />

Fax No. : (632) 927-7134<br />

E-mail : mrcampanano@yahoo.com<br />

Authorized Representative : Ms. Mirna R. Campañano<br />

Chief<br />

Accreditation No. Field <strong>of</strong> Testing<br />

LA-2010-177A Calibration<br />

Validity: 23 August 2015<br />

Products/<br />

Cl<strong>as</strong>s <strong>of</strong> Test<br />

Me<strong>as</strong>ured<br />

quantities/<br />

instrument<br />

Range to be<br />

calibrated<br />

195 <strong>of</strong> 391<br />

Best<br />

me<strong>as</strong>urement<br />

capability<br />

expressed <strong>as</strong> an<br />

uncertainty ()*<br />

Standard<br />

Method/<br />

Reference<br />

Standard<br />

5.03 Electrical Indicating and Recording Instrument<br />

.01 DC Voltmeters DC Voltmeters 0 to 1000V ±0.01% WI-CIC-26-VM /<br />

Fluke 5720A<br />

Calibrator<br />

.02 AC Voltmeters AC Voltmeters 0 to 750 V ±0.01% WI-CIC-26-VM /<br />

Fluke 5720A<br />

Calibrator<br />

.03 DC Ammeters DC Ammeters 0 to 2.2 A ±0.01% WI-CIC-27-AM /<br />

2.2 A to 3.2 A ±0.07% Fluke 5720A and<br />

3.2 A to 20 A ±0.1% Fluke 9100<br />

Calibrator<br />

.04 AC Ammeters AC Ammeters 0 to 2.2 A ±0.02% WI-CIC-27-AM /<br />

2.2 A to 3.2 A ±0.4% Fluke 5720A and<br />

3.2 A to 20 A ±0.3% Fluke 9100<br />

Calibrator<br />

.09 Ohmmeter Ohmmeter 0 to 200Ω ±0.6% WI-CIC-28-OM /<br />

200 Ω to 1 MΩ ±0.006% Fluke 5720A<br />

1 MΩ to 100 MΩ ±0.01% Calibrator<br />

.17 Digital Multimeter Digital Multimeter 0 to 1000V DC ±0.01% WI-CIC-19-FM /<br />

up to 4 ½ digits up to 4 ½ digits 0 to 750 V AC ±0.01% Fluke 5720A and<br />

0 to 2.2 A DC ±0.01% Fluke 9100<br />

2.2 A to 3.2 A DC ±0.07% Calibrator<br />

3.2 A to 20 A DC ±0.01%<br />

0 to 2.2 A AC ±0.02%<br />

2.2 A to 3.2 A AC ±0.4%<br />

3.2 A to 20 A AC ±0.3%<br />

0 to 200Ω ±0.6%<br />

200 Ω to 1 MΩ ±0.006%<br />

1 MΩ to 100 MΩ ±0.01%<br />

* A reported uncertainty will be that for the instrument itself during calibration plus the appropriate<br />

me<strong>as</strong>urement capability <strong>of</strong> the laboratory. The uncertainties are b<strong>as</strong>ed on an estimated confidence<br />

probability <strong>of</strong> approximate 95% unless otherwise stated.