Products/<br />

Cl<strong>as</strong>s <strong>of</strong> Test<br />

Me<strong>as</strong>ured<br />

quantities/<br />

instrument<br />

Range to be<br />

calibrated<br />

262 <strong>of</strong> 391<br />

Best<br />

me<strong>as</strong>urement<br />

capability<br />

expressed <strong>as</strong> an<br />

uncertainty ()*<br />

Standard Method/<br />

Reference<br />

Standard<br />

.04 A.C. ammeters AC Ammeter 0 to 400 µA 0.09 µA<br />

(from 50 Hz to 1<br />

kHz)<br />

0 to 4000 µA 0.9 µA<br />

0 to 40 mA 0.020 mA<br />

0 to 400 mA 0.14 mA<br />

0 to 4000 m A 1.5 mA<br />

0 to 10 A 0.006 A<br />

.09 Ohmmeters Ohmmeter 0 to 400 Ω 0.06 Ω<br />

0 to 40 kΩ 0.006 kΩ<br />

0 to 4 MΩ 0.000 7 MΩ<br />

0 to 40 MΩ 0.043 MΩ<br />

.17 Other<br />

Digital Multimeter<br />

EA-10/15: EA<br />

instruments up to 4 ½ digits<br />

only<br />

Guidelines<br />

@ DC voltage 0 to 400 mV 0.06 mV on the Calibration<br />

0 to 4 V 0.000 6 V <strong>of</strong> Digital<br />

0 to 40 V 0.006 V Multimeters<br />

0 to 400 V 0.06 V<br />

0 to 1000 V 0.6 V<br />

@ AC voltage 0 to 400 mV 0.06 mV<br />

(from 50 Hz to 1<br />

kHz)<br />

0 to 4 V 0.000 6 V<br />

0 to 40 V 0.006 V<br />

0 to 400 V 0.06 V<br />

0 to 1000 V 0.6 V<br />

@ DC current 0 to 400 µA 0.08 µA<br />

0 to 4000 µA 0.9 µA<br />

0 to 40 mA 0.006 mA<br />

0 to 400 mA 0.06 mA<br />

0 to 4000 m A 0.6 mA<br />

0 to 10 A 0.006 A<br />

@ AC current 0 to 400 µA 0.09 µA<br />

(from 50 Hz to 1<br />

kHz)<br />

0 to 4000 µA 0.9 µA<br />

0 to 40 mA 0.020 mA<br />

0 to 400 mA 0.14 mA<br />

0 to 4000 m A 1.5 mA<br />

0 to 10 A 0.006 A<br />

@ Resistance 0 to 400 Ω 0.06 Ω<br />

0 to 40 kΩ 0.006 kΩ<br />

0 to 4 MΩ 0.000 7 MΩ<br />

0 to 40 MΩ 0.043 MΩ<br />

5.45 Calibration <strong>of</strong> ancillary temperature me<strong>as</strong>uring instruments<br />

.05 Transmitters Temperature - 20 to 200 °C Input 0.2 °C ISA Textbook:<br />

Transmitters<br />

Output 0.006 mA Calibration –<br />

A Technician’s<br />

Guide.<br />

by Mike Cable<br />

* A reported uncertainty will be that for the instrument itself during calibration plus the appropriate<br />

me<strong>as</strong>urement capability <strong>of</strong> the laboratory. The uncertainties are b<strong>as</strong>ed on an estimated confidence<br />

probability <strong>of</strong> approximately 95% unless otherwise stated.<br />

Legends to Reference Standards:<br />

ASME - American Society <strong>of</strong> Mechanical Engineers<br />

ASTM - American Society for Testing and Materials<br />

BS - British Standard