Products/<br />

Cl<strong>as</strong>s <strong>of</strong> Test<br />

Me<strong>as</strong>ured<br />

quantities/<br />

instrument<br />

Range to be calibrated<br />

275 <strong>of</strong> 391<br />

Best<br />

me<strong>as</strong>urement<br />

capability<br />

expressed <strong>as</strong> an<br />

uncertainty ()*<br />

Standard Method/<br />

Reference<br />

Standard<br />

0 to 100 mV @ 50 kHz 0.043 mV<br />

100 mV to 1 V @ 1 kHz 0.000 10 V On-Site Calibration<br />

100 mV to 1 V @ 50 kHz 0.000 11 V for DC voltmeters<br />

1 V to 10 V @ 10 Hz 0.000 90 V and ohmmeters only<br />

1 V to 10 V @ 50 kHz 0.001 6 V<br />

10 V to 100 V @ 1 kHz 0.011 V<br />

10 V to 100 V @ 50 kHz 0.017 V<br />

100 V to 750 V @ 1 kHz 0.097 V<br />

.03 DC ammeters DC Ammeters 0 to 10 mA 0.000 68 mA<br />

10 mA to 100 mA 0.009 4 mA<br />

100 mA to 1 A 0.000 14 A<br />

1 A to 3 A 0.000 42 A<br />

.04 AC ammeters AC Ammeters 0 to 1 A 0.000 55 A<br />

1 to 3 A 0.001 A<br />

.09 Ohmmeters Resistance 0 to 100 Ω 0.001 8 Ω<br />

100 Ω to 1 kΩ 0.000 015 kΩ<br />

1 kΩ to 10 kΩ 0.000 15 kΩ<br />

10 kΩ to 100 kΩ 0.001 8 kΩ<br />

100 kΩ to 1 MΩ 0.000 03 MΩ<br />

1 MΩ to 10 MΩ 0.000 62 MΩ<br />

10 MΩ to 100 MΩ 0.019 MΩ<br />

5.37 Frequency and Time Me<strong>as</strong>uring Instruments and Standards<br />

.01 Frequency<br />

Meters<br />

Frequency Meters 0 to 100 mV @ 10 mV<br />

100 Hz<br />

100 mV to 1 V @ 1 V<br />

100 kHz<br />

5.42 High Voltage Equipment<br />

.01 Direct Voltage<br />

Test<br />

DC Voltage 0 to 5 kV<br />

.02 Alternating<br />

Voltage Test<br />

0.002 Hz<br />

0.002 kHz<br />

0.021 kV<br />

Non-Std. Method<br />

CD-WI-026<br />

Non-Std. Method<br />

CD-WI-032<br />

On-Site Calibration<br />

AC Voltage 0 to 5 kV 0.021 kV Non-Std. Method<br />

Cut-Off Current 0 to 10 mA 0.061 mA<br />

5.45 Calibration <strong>of</strong> ancillary temperature me<strong>as</strong>uring instruments<br />

.04 Indicators, Thermocouples - 250 to 1 700 °C<br />

recorders and type K, J, T, E, R<br />

controllers RTD type Pt100<br />

(385)<br />

- 200 to 400 °C<br />

0.31 °C<br />

0.17 °C<br />

CD-WI-032<br />

On-Site Calibration<br />

Non-Std. Method<br />

CD-WI-005<br />

On-Site Calibration<br />

* A reported uncertainty will be that for the instrument itself during calibration plus the appropriate<br />

me<strong>as</strong>urement capability <strong>of</strong> the laboratory. The uncertainties are b<strong>as</strong>ed on an estimated confidence<br />

probability <strong>of</strong> approximate 95% unless otherwise stated.