Products/<br />

Cl<strong>as</strong>s <strong>of</strong> Test<br />

Me<strong>as</strong>ured<br />

quantities/<br />

instrument<br />

Range to be<br />

calibrated<br />

388 <strong>of</strong> 391<br />

Best me<strong>as</strong>urement<br />

capability<br />

expressed <strong>as</strong> an<br />

uncertainty ()*<br />

Standard Method/<br />

Reference<br />

Standard<br />

Simulators by T/C Type T -200 to 390 C ±0.24 C VD-07-008<br />

Electrical T/C Type S 0 to 1700 C ±0.25 C<br />

Simulation) T/C Type R 0 to 1700 C ±0.29 C<br />

T/C Type E -200 to 990 C ±0.24 C<br />

T/C Type B 600 to 1800 C ±0.25 C<br />

T/C Type N -200 to 1290 C ±0.24 C<br />

T/C Type L -190 to 850 C ±0.24 C<br />

RTD Pt100 @385 -200 to 800 C ±0.24 C<br />

Generate:<br />

T/C Type K -200 to 1370 C ±0.32 C<br />

T/C Type J -200 to 1200 C ±0.27 C<br />

T/C Type T -200 to 400 C ±0.27 C<br />

T/C Type S 0 to 1700 C ±0.44 C<br />

T/C Type R 0 to 1700 C ±0.38 C<br />

T/C Type E -200 to 1000 C ±0.27 C<br />

T/C Type B 600 to 1800 C ±0.38 C<br />

T/C Type N -200 to 1290 C ±0.32 C<br />

T/C Type L -200 to 850 C ±0.32 C<br />

RTD Pt100 @385 -200 to 800 C ±0.01 C<br />

5.47 Testing <strong>of</strong> temperature controlled enclosures<br />

.01 Oven, Furnaces<br />

and Baths<br />

Calibration <strong>of</strong> Dry<br />

Block/Bath<br />

Calibrators<br />

Calibration <strong>of</strong><br />

Furnaces and Ovens<br />

-20 to 600 C ± 1 C In-house calibration<br />

QMP-07-T-012<br />

-20 to 250 C ± 1 C In-house/ Onsite<br />

calibration<br />

>250 to 600 C ± 2 C QMP-07-T-010<br />

VD-07-010<br />

0 to 200 °C ± 1 C In-house/ Onsite<br />

.02 Incubators Calibration <strong>of</strong><br />

Incubators<br />

calibration<br />

.03 Autoclaves and Calibration <strong>of</strong><br />

0 to 200 °C ± 1 C QMP-07-T-010<br />

sterilizing ovens Autoclaves and<br />

Sterilizing Ovens<br />

VD-07-010<br />

.04 Industrial Calibration <strong>of</strong><br />

- 20 to ambient ± 1 C In-house/ Onsite<br />

Freezers Industrial Freezers<br />

calibration<br />

.06 Other Calibration <strong>of</strong> Other -20 to 250 C ± 1 C QMP-07-T-010<br />

Enclosures<br />

5.48 Hygrometry<br />

Enclosures >250 to 600 C ± 2 C VD-07-010<br />

.01 Calibration <strong>of</strong><br />

humidity<br />

Calibration <strong>of</strong><br />

humidity me<strong>as</strong>uring<br />

-20 to 100 C ± 1 C<br />

In-house / Onsite<br />

calibration<br />

me<strong>as</strong>uring<br />

devices<br />

devices using<br />

chamber<br />

45 to 70% RH ±3.4 %RH<br />

QMP-07-T-009<br />

VD-07-011<br />

Calibration <strong>of</strong><br />

15 to 25 C ± 0.4 C In-house / Onsite<br />

humidity me<strong>as</strong>uring<br />

calibration<br />

devices using<br />

45 to 70% RH ±2.4 %RH QMP-07-T-009<br />

ambient room<br />

condition<br />

VD-07-011<br />

* A reported uncertainty will be that for the instrument itself during calibration plus the appropriate<br />

me<strong>as</strong>urement capability <strong>of</strong> the laboratory. The uncertainties are b<strong>as</strong>ed on an estimated confidence<br />

probability <strong>of</strong> approximately 95% unless otherwise stated.