Products/<br />

Cl<strong>as</strong>s <strong>of</strong> Test<br />

Me<strong>as</strong>ured<br />

quantities/<br />

instrument<br />

Range to be<br />

calibrated<br />

Page 7 <strong>of</strong> 391<br />

Best<br />

me<strong>as</strong>urement<br />

capability<br />

expressed <strong>as</strong> an<br />

uncertainty ()*<br />

Standard Method/<br />

Reference Standard<br />

.17 Other<br />

Calibration <strong>of</strong> Wrist 0 to 100 MΩ ± 3 %<br />

TR-WI-RES-002<br />

In-house calibration/<br />

Instruments Strap / Foot Strap<br />

On-site calibration<br />

Checkers<br />

TR-WI-RES-003<br />

Calibration <strong>of</strong><br />

0 to 100 MΩ ± 3 % In-house calibration/<br />

Insulation Testers<br />

On-site calibration<br />

TR-WI-RES-004<br />

Calibration <strong>of</strong><br />

0 to 1000 V DC<br />

± 0.01 % In-house calibration/<br />

Digital Multimeters 0 to 750 V AC @ 60 ± 0.5 % On-site calibration<br />

up to 4½ digits only<br />

Hz<br />

TR-WI-DMM-001<br />

0 to 10 A DC<br />

± 0.1 %<br />

0 to 3 A AC @ 60 Hz<br />

0 to 100 MΩ<br />

± 0.5 %<br />

± 3 %<br />

5.37 Frequency and Time Me<strong>as</strong>uring Instruments and Standards<br />

.01 Frequency Frequency Meters 10 Hz to 100 Hz ± 0.06 % In-house calibration<br />

Meters<br />

100 Hz to 10 MHz ± 0.006 % TR-WI-FRE-003<br />

.03 Counters Frequency<br />

10 Hz to 100 Hz ± 0.06 % In-house calibration<br />

Counters<br />

100 Hz to 10 MHz ± 0.006 % TR-WI-FRE-003<br />

.05 Clocks and Calibration <strong>of</strong><br />

5 sec. to 30 min. ± 0.5 sec. In-house calibration/<br />

Watches Timers<br />

On-site calibration<br />

TR-WI-FRE-004<br />

5.38 Waveform Me<strong>as</strong>uring Instruments<br />

.05 Other<br />

Calibration <strong>of</strong><br />

1 mV to 100 V ± 3 % amplitude In-house calibration/<br />

Characteristics Oscilloscopes<br />

5 nsec. to 5 sec. ± 1 % time mark On-site calibration<br />

@ Vertical and<br />

Horizontal<br />

Deflections<br />

TR-WI-FRE-001<br />

5.39 Power Supplies<br />

.01 Power<br />

Calibration <strong>of</strong> DC 0 to 1 000 V DC ± 0.01 % In-house calibration/<br />

Supplies<br />

Power Supplies<br />

0 to 10 A DC<br />

± 0.1 % On-site calibration<br />

TR-WI-POW-001<br />

5.42 High Voltage Equipment<br />

.01 Direct Voltage Calibration <strong>of</strong> High 0 to 1 kV ± 3 % In-house calibration/<br />

Test<br />

Voltage DC<br />

On-site calibration<br />

.02 Alternating Calibration <strong>of</strong> High 0 to 5 kV ± 3 % TR-WI-HIV-001<br />

Voltage Test Voltage AC<br />

.06 Dielectric Test Calibration <strong>of</strong><br />

Dielectric Voltage<br />

AC<br />

0 to 5 kV ± 3 %<br />

5.44 Calibration <strong>of</strong> Temperature Me<strong>as</strong>urement Equipment<br />

.02 B<strong>as</strong>e Metal Calibration <strong>of</strong> B<strong>as</strong>e 0 to 95 °C ± 0.5 °C In-house calibration/<br />

Metal<br />

On-site calibration<br />

Thermocouples Thermocouples<br />

TR-WI-TEM-002<br />

.04 Metallic Calibration <strong>of</strong> RTD 0 to 95 °C ± 0.5 °C<br />

Resistance and Other Metallic<br />

Thermometers Resistance<br />

Thermometers<br />

5.45 Calibration <strong>of</strong> Ancillary Temperature Me<strong>as</strong>uring Instruments<br />

.04 Indicators, Calibration <strong>of</strong> - 200 to 1 350 °C ± 1 °C In-house calibration/<br />

Recorders and Temperature<br />

On-site calibration<br />

Controllers Indicators,<br />

Controllers and<br />

Recorders<br />

TR-WI-TEM-001<br />

5.47 Testing <strong>of</strong> Temperature Controlled Enclosures<br />

.01 Ovens, Calibration <strong>of</strong> Ambient to 500 °C ± 2 °C In-house calibration/<br />

Furnaces and Ovens<br />

On-site calibration<br />

Baths Calibration <strong>of</strong><br />

Furnaces<br />

Ambient to 800 °C ± 3 °C<br />

TR-WI-TEM-003<br />

Calibration <strong>of</strong> Baths Ambient to 500 °C ± 2 °C<br />

.02 Incubators Calibration <strong>of</strong><br />

Incubators<br />

Ambient to 80 °C ± 2 °C