Cl<strong>as</strong>s <strong>of</strong> Test<br />

Me<strong>as</strong>ured<br />

quantities/<br />

instrument<br />

Range to be<br />

calibrated<br />

386 <strong>of</strong> 391<br />

Best me<strong>as</strong>urement<br />

capability<br />

expressed <strong>as</strong> an<br />

uncertainty ()*<br />

Standard Method/<br />

Reference<br />

Standard<br />

5.35 Indicating and recording instruments<br />

.01 D.C. voltmeters Calibration <strong>of</strong> DC 0 to 300 mV ±0.003% FS In-house calibration<br />

Voltage Me<strong>as</strong>ure 0 to 3 V ±0.0004% FS QMP-07-E-001<br />

0 to 30 V ±0.0003% FS<br />

0 to 300 V ±0.0003% FS<br />

0 to 1000 V ±0.0001% FS<br />

.02 A.C. voltmeters Calibration <strong>of</strong> AC 0 to 300 mV ±0.03% FS In-house calibration<br />

Voltage Me<strong>as</strong>ure 0 to 3 V ±0.004% FS QMP-07-E-001<br />

0 to 30 V ±0.002% FS<br />

0 to 100 V ±0.006% FS<br />

0 to 300 V ±0.002% FS<br />

0 to 500 V ±0.01% FS<br />

0 to 750 V ±0.008% FS<br />

.03 D.C. ammeters Calibration <strong>of</strong> DC 0 to 300 µA ±0.02% FS In-house calibration<br />

Current Me<strong>as</strong>ure 0 to 3 mA ±0.003% FS QMP-07-E-002<br />

0 to 30 mA ±0.002% FS<br />

0 to 300 mA ±0.0003% FS<br />

0 to 3 A ±0.002% FS<br />

0 to 10 A ±0.006% FS<br />

.04 A.C. ammeters Calibration <strong>of</strong> AC 0 to 300 µA ±0.08% FS In-house calibration<br />

Current Me<strong>as</strong>ure 0 to 3 mA ±0.008% FS QMP-07-E-002<br />

0 to 30 mA ±0.02% FS<br />

0 to 300 mA ±0.002% FS<br />

0 to 3 A ±0.002% FS<br />

0 to 10 A ±0.06% FS<br />

.09 Ohmmeters Calibration <strong>of</strong> 0 to 40 ±0.03% FS In-house calibration<br />

Resistance Me<strong>as</strong>ure 0 to 400 ±0.002% FS QMP-07-E-003<br />

0 to 4 k ±0.0003% FS<br />

0 to 40 k ±0.0002% FS<br />

0 to 400 k ±0.0002% FS<br />

0 to 4 M ±0.001% FS<br />

0 to 40 M ±0.0002% FS<br />

0 to 100 M ±0.006% FS<br />

.10 LCR Meters Calibration <strong>of</strong> Low 0.5 to 4 nF ± 0.20 % FS In-house calibration<br />

Capacitance Meters > 4 to 40 nF ± 0.16 % FS QMP-07-E-004<br />

> 40 to 400 nF ± 0.08 % FS<br />

> 400nF to 4 µF ± 0.16 % FS<br />

> 4 to 40 µF ± 0.16 % FS<br />

> 40 to 400 µF ± 0.16 % FS<br />

> 400 µF to 4 mF ± 0.15 % FS<br />

.17 Other<br />

instruments<br />

Calibration <strong>of</strong> 4 ½<br />

digits Digital<br />

Multimeters<br />

> 4 to 30 mF ± 0.1 % FS<br />

@ DC Voltage 0 to 300 mV ±0.003% FS<br />

0 to 3 V ±0.0004% FS<br />

0 to 30 V ±0.0003% FS<br />

0 to 300 V ±0.0003% FS<br />

0 to 1000 V ±0.0001% FS<br />

@ AC Voltage, 60Hz 0 to 300 mV ±0.03% FS<br />

0 to 3 V ±0.004% FS<br />

0 to 30 V ±0.002% FS<br />

0 to 100 V ±0.006% FS<br />

0 to 300 V ±0.002% FS<br />

0 to 500 V ±0.01% FS<br />

0 to 750 V ±0.008% FS<br />

@ DC Current 0 to 300 µA ±0.02% FS<br />

0 to 3 mA ±0.003% FS<br />

0 to 30 mA ±0.002% FS<br />

0 to 300 mA ±0.0003% FS<br />

0 to 3 A ±0.002% FS<br />

0 to 10 A ±0.006% FS<br />

In-house calibration<br />

QMP-07-E-018