13.2 Technical specifications<br />

Product<br />

Manufacturer and distributor<br />

Classification<br />

Overall system (excluding pump)<br />

Electro-pneumatic ventricular assist <strong>Ikus</strong> EXCOR VAD with<br />

EXCOR stationary Driving Unit <strong>Ikus</strong><br />

<strong>Berlin</strong> <strong>Heart</strong> GmbH<br />

Wiesenweg 10<br />

12247 <strong>Berlin</strong><br />

Germany<br />

Class 3<br />

Ambient operating temperature + 10 to +30 °C, with restrictions up to 35 °C<br />

Ambient temperature in storage and<br />

transport<br />

Max. ambient magnetic field 10 A/m<br />

Relative ambient air humidity 45 to 75%<br />

Ambient atmospheric pressure 900 hPa to 1060 hPa<br />

Pump<br />

Dimensions see product datasheets<br />

- 10 to +50 °C<br />

6 h rest phase after transport be<strong>for</strong>e commissioning<br />

Material Housing and membranes: polyurethane<br />

Driving tube adapter: polyoxymethylene<br />

Connectors: titanium<br />

Coating of the blood contact surfaces Carmeda ® BioActive Surface (CBAS TM )<br />

Ambient temperature of blood pump storage max. +30 °C<br />

Duration of <strong>use</strong> recommended: 1 year; max.: 500 days<br />

Cannulas<br />

Dimensions see product datasheets<br />

Material Silicone; partially textile-rein<strong>for</strong>ced; partially surrounded<br />

by suturable polyester velour; with partial flexible metal<br />

rein<strong>for</strong>cement: wire 2 mm, steel Rd 1.4301; apex cannula<br />

with titanium sheath; mandrin POM<br />

<strong>Ikus</strong><br />

Dimensions (WxHxD) 46 x 95 x 73 cm (with the laptop closed)<br />

46 x 120 x 73 cm (with the laptop open)<br />

Driving tube (red/blue) Recommended lifespan: 1 year; max.: 500 days<br />

Weight 93 kg<br />

Input voltage AC 115 V or AC 230 V (see identification plate)<br />

Power input 575 VA<br />

Mains power f<strong>use</strong> 5 A (115 V) to 3 A (230 V)<br />

External alarm connector Electrical data: max. 24 V /1 A<br />

Insulation data: 2.5mm/ 4 mm strike and creep distance between<br />

alarm contact and 24V safety-low voltage inside the device<br />

(coil side)<br />

Insulation test voltage: 500V<br />

Protection class IPX0 (protection against accidental contact not tested, no water<br />

protection)<br />

Pump frequency 30 to 150 bpm<br />

Systolic pressure 60 to 350 mmHg<br />

Technical specifications<br />

1000002 Rev. 4.9.1 89