Batteries discharged; battery mode not possible<br />
Error messages and corrective action<br />
incomplete. There<strong>for</strong>e, only end the monitoring program when it is absolutely necessary<br />
(<strong>for</strong> example, to set new <strong>use</strong>r profiles)! Then restart it again as soon as possible!<br />
11.1 Batteries discharged; battery mode not possible<br />
The batteries have become discharged during running mains power operation.<br />
Warning!<br />
• You must continue to operate the <strong>Ikus</strong> on mains power! There is a threat of total failure<br />
if battery mode is <strong>use</strong>d!<br />
<strong>Instructions</strong><br />
1. Check the status of the patient.<br />
2. Check the filling and emptying of the blood pump(s), and monitor the function of the<br />
<strong>Ikus</strong>. Do not operate the <strong>Ikus</strong> without supervision!<br />
3. If possible: provide the patient with a replacement <strong>Ikus</strong> and switch off the malfunctioning<br />
<strong>Ikus</strong> (see Section 12.4, p. 80). If no replacement <strong>Ikus</strong> is available: Supply the patient, if<br />
necessary (in the event of a failure malfunction of the defective <strong>Ikus</strong>), by means of the<br />
manual pump (see Section 12.5, p. 81)!<br />
4. Always do the following immediately:<br />
Notify hotline! +49(0)30 8187 2772<br />
11.2 Updating the parameter block failed<br />
The <strong>Ikus</strong> cannot store the changed parameter values in the internal system memory. The<br />
driving unit continues to work with the changed values; however, in the case of an<br />
internal reset triggered by the <strong>Ikus</strong>, the old values would become valid again.<br />
Notify hotline! +49(0)30 8187 2772<br />
<strong>Instructions</strong><br />
1. Check the parameter values regularly. After a reset, if the <strong>Ikus</strong> continues to work with the<br />
old parameter values: Readjust the parameter values.<br />
Note<br />
• During a reset, the <strong>Ikus</strong> emits a short acoustic alarm.<br />
11.3 The acoustic alarm is not correctly detected<br />
If, within 8 seconds, the message Acoustic alarm: OK appears,<br />
the <strong>Ikus</strong> is working perfectly. No further corrective action is necessary.<br />
If the message Acoustic alarm: OK does not appear within 8 seconds,<br />
the alarm circuit is defective. A possible error might not be detected.<br />
<strong>Instructions</strong><br />
1. Check the status of the patient.<br />
2. Check the filling and emptying of the blood pump(s), and monitor the function of the<br />
<strong>Ikus</strong>. Important: Do not operate the <strong>Ikus</strong> without supervision!<br />
3. If possible,: provide the patient with a replacement <strong>Ikus</strong> and switch off the malfunctioning<br />
<strong>Ikus</strong> (see Section 12.4, p. 80). If no replacement <strong>Ikus</strong> is available: Supply the patient, if<br />
necessary (in the event of a failure malfunction of the defective <strong>Ikus</strong>), by means of the<br />
manual pump (see Section 12.5, p. 81)!<br />
4. Always do the following immediately:<br />
Notify hotline! +49(0)30 8187 2772<br />
64 1000002 Rev. 4.9.1