Batteries discharged; battery mode not possible<br />

Error messages and corrective action<br />

incomplete. There<strong>for</strong>e, only end the monitoring program when it is absolutely necessary<br />

(<strong>for</strong> example, to set new <strong>use</strong>r profiles)! Then restart it again as soon as possible!<br />

11.1 Batteries discharged; battery mode not possible<br />

The batteries have become discharged during running mains power operation.<br />

Warning!<br />

• You must continue to operate the <strong>Ikus</strong> on mains power! There is a threat of total failure<br />

if battery mode is <strong>use</strong>d!<br />

<strong>Instructions</strong><br />

1. Check the status of the patient.<br />

2. Check the filling and emptying of the blood pump(s), and monitor the function of the<br />

<strong>Ikus</strong>. Do not operate the <strong>Ikus</strong> without supervision!<br />

3. If possible: provide the patient with a replacement <strong>Ikus</strong> and switch off the malfunctioning<br />

<strong>Ikus</strong> (see Section 12.4, p. 80). If no replacement <strong>Ikus</strong> is available: Supply the patient, if<br />

necessary (in the event of a failure malfunction of the defective <strong>Ikus</strong>), by means of the<br />

manual pump (see Section 12.5, p. 81)!<br />

4. Always do the following immediately:<br />

Notify hotline! +49(0)30 8187 2772<br />

11.2 Updating the parameter block failed<br />

The <strong>Ikus</strong> cannot store the changed parameter values in the internal system memory. The<br />

driving unit continues to work with the changed values; however, in the case of an<br />

internal reset triggered by the <strong>Ikus</strong>, the old values would become valid again.<br />

Notify hotline! +49(0)30 8187 2772<br />

<strong>Instructions</strong><br />

1. Check the parameter values regularly. After a reset, if the <strong>Ikus</strong> continues to work with the<br />

old parameter values: Readjust the parameter values.<br />

Note<br />

• During a reset, the <strong>Ikus</strong> emits a short acoustic alarm.<br />

11.3 The acoustic alarm is not correctly detected<br />

If, within 8 seconds, the message Acoustic alarm: OK appears,<br />

the <strong>Ikus</strong> is working perfectly. No further corrective action is necessary.<br />

If the message Acoustic alarm: OK does not appear within 8 seconds,<br />

the alarm circuit is defective. A possible error might not be detected.<br />

<strong>Instructions</strong><br />

1. Check the status of the patient.<br />

2. Check the filling and emptying of the blood pump(s), and monitor the function of the<br />

<strong>Ikus</strong>. Important: Do not operate the <strong>Ikus</strong> without supervision!<br />

3. If possible,: provide the patient with a replacement <strong>Ikus</strong> and switch off the malfunctioning<br />

<strong>Ikus</strong> (see Section 12.4, p. 80). If no replacement <strong>Ikus</strong> is available: Supply the patient, if<br />

necessary (in the event of a failure malfunction of the defective <strong>Ikus</strong>), by means of the<br />

manual pump (see Section 12.5, p. 81)!<br />

4. Always do the following immediately:<br />

Notify hotline! +49(0)30 8187 2772<br />

64 1000002 Rev. 4.9.1