11.27.3 Additional messages during the start test<br />

�<br />

Changes:<br />

• The section has been supplemented.<br />

• Error: 0000 0011 0011 0111 (fault in power supply)<br />

Replacing the blood pump<br />

12.1 Replacing the blood pump(s)<br />

Changes:<br />

• This section has been supplemented by a note on safety.<br />

• This section has been supplemented by section 12.1.1 Replacing blood pump<br />

beca<strong>use</strong> of growth of the patient.<br />

WARNING<br />

CAUTION<br />

HOTLINE<br />

Remove cable tie at the connector stub of the blood pump with a blunt<br />

tool. IMPORTANT: do not <strong>use</strong> pointed or sharp-edged objects (surgical<br />

instruments) under any circumstances. Otherwise the cannula may be<br />

damaged.<br />

12.1.1 Replacing the blood pump beca<strong>use</strong> of growth of the patient<br />

For children, plan <strong>for</strong> a switch to a larger blood pump(s) ahead of time.<br />

Otherwise there is a danger of inadequate support due to a discharge<br />

rate that is too low.<br />

It is possible that the blood pump selected at the time of implantation cannot<br />

guarantee adequate support <strong>for</strong> the entire period of therapy. Physical growth and/or<br />

an increase in weight of the patient can mean that the patient is not adequately<br />

supported with the present pump. There<strong>for</strong>e plan <strong>for</strong> a replacement of the blood<br />

pump ahead of time. Refer to the table showing the ratio between body weight and<br />

pump size in section 13.1 Overview: product range and combination options. This<br />

table serves solely <strong>for</strong> orientation purposes and cannot be applied in every individual<br />

case. The surgeon must select a pump in consultation with the <strong>Berlin</strong> <strong>Heart</strong> GmbH.<br />

Notify hotline! +49 (0)30 81 87 27 72<br />

When replacing the blood pump(s), proceed according to the instructions <strong>for</strong> <strong>use</strong>.<br />

See section 12.1 of the instructions <strong>for</strong> <strong>use</strong>.<br />

1000095x14 Rev. 8 25/28