Intraoperative drive management<br />

Start up the <strong>Ikus</strong> and set the parameters<br />

remains identical.<br />

• separate pulse is sometimes helpful in patients with intracardial shunts.<br />

<strong>Instructions</strong><br />

1. Move the cursor with ← / → to the field with the current operating mode display.<br />

A popup menu appears showing the possible operating modes.<br />

2. Select the desired operating mode with ↓,↑, then confirm with . The system<br />

works in the set operating mode.<br />

Target values<br />

Systole [mmHg]<br />

left/right<br />

The most important criterion when selecting the drive parameters is good filling and emptying<br />

of the pump; the settings must be oriented toward achieving this goal.<br />

Note<br />

• The systolic driving pressure must be higher than the systolic pressure of the patient.<br />

Important: When the systolic duration is shortened or with very small cannulas, a higher<br />

value may need to be selected than recommended here.<br />

• The driving pressures reached are dependent on the diameter of the cannulas <strong>use</strong>d.<br />

• The following values are only guideline values; they are not authoritative in individual<br />

cases!<br />

Diastole [mmHg]<br />

left/right<br />

Recommended target values <strong>for</strong> normal operation<br />

Recommendation<br />

• Now remove the de-airing needle (see page 51).<br />

Warning!<br />

• On removing the de-airing needle, never pull on the de-airing tube, but rather on the deairing<br />

needle itself.<br />

6.2.9 Switching from HLM to VAD<br />

Rate [bpm]<br />

220/150 -40/-40 80 40/40<br />

rel. systolic duration [%]<br />

left/right<br />

The goal is to reduce the HLM flow and thus shift the volume from the HLM to the<br />

patient (to the VAD).<br />

<strong>Instructions</strong><br />

1. If firstly the right blood pump and then the left blood pump too begin to fill, reduce the<br />

HLM flow and increase the EXCOR rate firstly from 30 bpm stepwise, until the desired<br />

flow is reached. Important: In doing so, ensure sufficient filling, and if necessary regulate<br />

the driving pressure!<br />

2. If necessary, regulate the suction pressure.<br />

Note<br />

• If stage cannulas or a connecting set are <strong>use</strong>d, the pumps do not eject fully at a rate<br />

of > 100 bpm. With these cannulas, rates of > 100 bpm should be avoided.<br />

Recommendation<br />

• To secure the driving tubes to the cannulas, as soon as the system is running stably<br />

(see page 51).<br />

38 1000002 Rev. 4.9.1