EXCOR cannulas<br />

Description: Blood pump, cannulas, accessories<br />

EXCOR blood pumps consist of a transparent polyurethane (PU) housing divided by a<br />

flexible triple-layer membrane into a blood chamber and an air chamber.<br />

From the blood chamber, an inflow or outflow stub lead to an inflow or outflow cannula.<br />

The stubs themselves consist of PU. At their ends, there are titanium connectors to<br />

which the cannulas are attached. Mechanical valves in the stubs guarantee a targeted<br />

blood flow. EXCOR blood pumps are available with tilting-disk valves (50 - 80 ml stroke<br />

volume) or trileaflet valves made from polyurethane (10 - 80 ml stroke volume).<br />

The blood contact surface of the pump is coated with Carmeda ® BioActive Surface<br />

(CBASTM ). The transparent housing of the blood pump permits easy visual inspection,<br />

so that the filling and emptying and the status of the blood chamber can easily be<br />

monitored.<br />

By means of the de-airing nipple in the blood chamber, the pump is de-aired during<br />

commissioning with the aid of the de-airing set.<br />

The air chamber of the pump is equipped with a driving tube connector. This connects<br />

to the air-conducting driving tube that links the pump to the Driving Unit <strong>Ikus</strong>. With<br />

suction and driving pressures, the <strong>Ikus</strong> actuates the triple-layer membrane of the blood<br />

pump. To minimize friction, there is graphite powder between the individual<br />

membranes.<br />

3.2 EXCOR cannulas<br />

For EXCOR, three cannula types in three different sizes are offered:<br />

• Atrial cannulas (inflow cannulas)<br />

• Apex cannulas (inflow cannulas)<br />

• Arterial cannulas (outflow cannulas)<br />

The cannulas consist of tissue-friendly silicon. Suture rings made from polyester velour<br />

permit fast and safe anastomosis of the cannulas. The midsection of every cannula is<br />

sheathed with polyester velour to ensure good inward movement of the cannula.<br />

Some arterial cannulas possess a flexible metal rein<strong>for</strong>cement that permits the<br />

adjustment of the cannulas to the anatomical conditions.<br />

Some atrial cannulas are delivered with a mandrin.<br />

1 2 3<br />

Cannula heads: 1) Atrial cannula, 2) apex cannula, 3) arterial cannula<br />