EXCOR cannulas<br />
Description: Blood pump, cannulas, accessories<br />
EXCOR blood pumps consist of a transparent polyurethane (PU) housing divided by a<br />
flexible triple-layer membrane into a blood chamber and an air chamber.<br />
From the blood chamber, an inflow or outflow stub lead to an inflow or outflow cannula.<br />
The stubs themselves consist of PU. At their ends, there are titanium connectors to<br />
which the cannulas are attached. Mechanical valves in the stubs guarantee a targeted<br />
blood flow. EXCOR blood pumps are available with tilting-disk valves (50 - 80 ml stroke<br />
volume) or trileaflet valves made from polyurethane (10 - 80 ml stroke volume).<br />
The blood contact surface of the pump is coated with Carmeda ® BioActive Surface<br />
(CBASTM ). The transparent housing of the blood pump permits easy visual inspection,<br />
so that the filling and emptying and the status of the blood chamber can easily be<br />
monitored.<br />
By means of the de-airing nipple in the blood chamber, the pump is de-aired during<br />
commissioning with the aid of the de-airing set.<br />
The air chamber of the pump is equipped with a driving tube connector. This connects<br />
to the air-conducting driving tube that links the pump to the Driving Unit <strong>Ikus</strong>. With<br />
suction and driving pressures, the <strong>Ikus</strong> actuates the triple-layer membrane of the blood<br />
pump. To minimize friction, there is graphite powder between the individual<br />
membranes.<br />
3.2 EXCOR cannulas<br />
For EXCOR, three cannula types in three different sizes are offered:<br />
• Atrial cannulas (inflow cannulas)<br />
• Apex cannulas (inflow cannulas)<br />
• Arterial cannulas (outflow cannulas)<br />
The cannulas consist of tissue-friendly silicon. Suture rings made from polyester velour<br />
permit fast and safe anastomosis of the cannulas. The midsection of every cannula is<br />
sheathed with polyester velour to ensure good inward movement of the cannula.<br />
Some arterial cannulas possess a flexible metal rein<strong>for</strong>cement that permits the<br />
adjustment of the cannulas to the anatomical conditions.<br />
Some atrial cannulas are delivered with a mandrin.<br />
1 2 3<br />
Cannula heads: 1) Atrial cannula, 2) apex cannula, 3) arterial cannula<br />
10 1000002 Rev. 4.9.1