25.10.2012 Views

Instructions for use: Ikus - Berlin Heart

Instructions for use: Ikus - Berlin Heart

Instructions for use: Ikus - Berlin Heart

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

9.5.1 Visual inspection: Filling/ ejecting of the blood pump<br />

Regular inspection of blood pump(s) and cannulas<br />

A blood pump fills and empties optimally if, in the end-of-systole and end-of-diastole<br />

positions, the membrane surface is completely smooth. Check over several pump cycles<br />

that the pump(s) is/are filling and emptying completely. If a pump does not fill and/or<br />

empty completely, then corresponding corrective action must be taken.<br />

Corrective action<br />

• For each blood pump: Check the position and condition of the driving tube and cannulas.<br />

(Inflow deterioration due to breaking off of the cannulas/driving tubes is rather<br />

uncommon.)<br />

• For each blood pump: Check the membrane movement.<br />

• Medicinal monitoring of the patient: Check CVP; check mean arterial pressure and adjust<br />

if necessary.<br />

• Check volume situation:<br />

• Quantity of bleeding<br />

• Increased elimination (diuretics?)<br />

• Tamponade<br />

• Important: Increasing the negative pressure cannot ca<strong>use</strong> any major<br />

improvement if a corresponding volume supply is not available.<br />

• For LVAD: Pay attention to the right heart function.<br />

Adjusting the parameter values<br />

Only adjust the parameters if the above-mentioned corrective action has no effect or <strong>for</strong><br />

• Mobilization of patient: Adjustment of the left and right systolic pressure. Do not reduce<br />

pressures which have been increased while the patient is lying down.<br />

• Signs of cardiac insufficiency: With good membrane movement, a decline in urine<br />

elimination, lactate increase and shortness of breath must be observed. In this case,<br />

increase the rate and adjust other settings. If necessary, consider replacing the pump<br />

(larger volume).<br />

<strong>Instructions</strong><br />

1. Move the cursor with ← / → to the desired field in the parameter table. The active field<br />

is colored underneath.<br />

2. Adjust the value with ↓,↑/ / , then confirm with . The system<br />

works with the new value.<br />

Cautionary measure<br />

• Confirm each changed parameter value individually with . The system only<br />

applies the changed values after this confirmation.<br />

9.5.2 Visual inspection: Deposits<br />

At least three times per day, check the blood pump(s) <strong>for</strong> deposits (fibrin, thrombi).<br />

Record any findings in the “EXCOR Pump Log” (standard <strong>for</strong>m see page 94). If deposits<br />

are beginning: Check the anticoagulation therapy!<br />

1000002 Rev. 4.9.1 59

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!